Waldemar Link GmbH & Co. KG (Mfg Site): Endo Model Modular Femur and Tibia, PorEx. orthopedic prosthesis. Model Nos.: X-Small, Right 15-3815/11, Small, Right 15-3816/11, Small, Left 15-3816/12, Medium, Right 15-3817/11, Medium, Left 15-3817/12, Large, Right 15-3818/11.

Recall number
Z-1378-2022
Recalling firm
Waldemar Link GmbH & Co. KG (Mfg Site)
Product
Endo Model Modular Femur and Tibia, PorEx. orthopedic prosthesis. Model Nos.: X-Small, Right 15-3815/11, Small, Right 15-3816/11, Small, Left 15-3816/12, Medium, Right 15-3817/11, Medium, Left 15-3817/12, Large, Right 15-3818/11.
Status
Open, Classified
Initiated
2022-05-24
Posted
not on file
Terminated
not on file
Reason
There is a risk that blind screws of the modular tibial component cannot be loosened intraoperatively. This may lead to prolongation of surgery due to an intraoperative change in procedure.
Root cause
Nonconforming Material/Component
Action
On May 24, 2022, the firm notified affected customers via Urgent Field Safety Notice email. Customers were provided a modified surgical technique providing information for the cementing technique for Tilastan proximal spacers to minimize the risk. No product return is required.
Quantity
3785
Distribution
Domestic distribution to AL, CA, FL, GA, IL, IN, KS, MI, MN, NV, NY, OH, TN, TX, WA. Foreign distribution to Algeria Argentina Australia Austria Belgium Bulgaria Canada China Colombia Denmark Ecuador Estonia Finland France Germany Great Britain Greece Hungary India Indonesia Israel Italy Latvia Libya Lithuania Mexico Netherlands Norway Pakistan Peru Philippines Poland Romania Saudi Arabia Slovakia Slovenia Spain Sudan Sweden Switzerland Thailand Turkey Ukraine United Arab Emirates Uruguy Vietnam
Lot, serial or model codes
UDI-DI  04026575034727 X-Small, Right; 04026575034741 Small, Right; 04026575034758 Small, Left; 04026575164042 Medium, Right; 04026575164059 Medium, Left; 04026575164028 Large, Right;   Serial/Lot Numbers: 200818/1709 200818/1720 201013/0291 200513/2160 200818/1732 200623/4154 200812/1511 210302/2440
510(k) numbers
K212742
PMA numbers
not on file
Product code
JWH
Firm FEI
3003386935
Firm city
Norderstedt
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source