Skip to the content

FDA Medical Device Recalls (openFDA)

Cook Inc.: MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000 US) G46022 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.

Recall number
Z-1377-2018
Recalling firm
Cook Inc.
Product
MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000 US) G46022 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.
Status
Terminated
Initiated
2018-01-12
Posted
2018-04-13
Terminated
2020-05-08
Reason
The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.
Root cause
Device Design
Action
Cook plans to begin notifying consignees of the recall on January 12, 2018 Written recall communication letters will be mailed via courier to US consignees
Quantity
483
Distribution
Nationwide (K-IMNC-100 -US G46022) . (K-IMNC-100 G20079 manufactured in Australia and distributed worldwide)
Lot, serial or model codes
All lots
510(k) numbers
["K983642"]
PMA numbers
not on file
Product code
MQG
Firm FEI
1820334
Firm city
Bloomington
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register