FDA Medical Device Recalls (openFDA)
Baxter Healthcare Corporation: ABACUS TPN (Total Parenteral Nutrition) Calculation software, Product codes: 8300-0167 (Abacus V3.1 CE), 8300-0168 (Abacus V3.1 SE), 8300-0169 (Abacus V3.1 ME), 8300-0191 (Abacus V3.2 CE), 8300-0192 (Abacus V3.2 SE), 8300-0193 (Abacus V3.2 ME), 8300-3391 (Abacus V3.3 CE), 8300-3392 (Abacus V3.3 SE), and 8300-3393 (Abacus V3.3 ME). Commonly used with ExactaMix automated compounding devices (ExactaMix 1200 and ExactaMix 2400).
- Recall number
- Z-1376-2022
- Recalling firm
- Baxter Healthcare Corporation
- Product
- ABACUS TPN (Total Parenteral Nutrition) Calculation software, Product codes: 8300-0167 (Abacus V3.1 CE), 8300-0168 (Abacus V3.1 SE), 8300-0169 (Abacus V3.1 ME), 8300-0191 (Abacus V3.2 CE), 8300-0192 (Abacus V3.2 SE), 8300-0193 (Abacus V3.2 ME), 8300-3391 (Abacus V3.3 CE), 8300-3392 (Abacus V3.3 SE), and 8300-3393 (Abacus V3.3 ME). Commonly used with ExactaMix automated compounding devices (ExactaMix 1200 and ExactaMix 2400).
- Status
- Terminated
- Initiated
- 2022-06-22
- Posted
- 2022-07-22
- Terminated
- 2024-10-09
- Reason
- There is a potential risk of medication error when using specific Abacus Order Entry & Calculation software.
- Root cause
- Software design
- Action
- Baxter sent an Urgent Medical Device Correction notice on 06/22/2022 by letter. The notice explained the problem and the hazard and recommended the following: customers refrain from making updates to bag label templates and to contact Baxter Technical Support if updates to bag label templates are needed. Please continue to follow the instructions in the Abacus User and Abacus Configuration Guides when creating an order, and have a pharmacist thoroughly review all order outputs including the printed label for accuracy and safety. As mentioned above, please ensure your facility processes include pharmacy checks during the compounding process as well as a nursing check. Baxter will communicate to customers when the software upgrade becomes available. At that time additional instructions to obtain the software upgrade will be provided.
- Quantity
- 9 units
- Distribution
- United States (including Puerto Rico), Argentina, Bermuda, Chile, Colombia, Dominican Republic, Panama, Canada, Vietnam, Hong Kong, Singapore, and EMEA
- Lot, serial or model codes
- Product Code 8300-3391, UDI (01)05413765577345(10)3.3.2.1, Release date 20-Sep-2019; Product Code 8300-3392, UDI (01)05413765577352(10)3.3.2.1, Release date 20-Sep-2019; Product Code 8300-3393, UDI (01)05413765577369(10)3.3.2.1, Release date 20-Sep-2019; Product Codes: 8300-0167, V3.1 CE, Release date 20-May-2013, 8300-0168, V3.1 SE, Release date 20-May-2013, 8300-0169, V3.1 ME, Release date 20-May-2013, 8300-0191, V3.2 CE, Release date 17-Dec-2015, 8300-0192, V3.2 SE, Release date 17-Dec-2015, 8300-0193, V3.2 ME, Release date 17-Dec-2015, No UDIs
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- NEP
- Firm FEI
- 1417572
- Firm city
- Deerfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28