Terumo Medical Corporation: Portico Solo Re-Collapsible Access System , Model Numbers PRTSOLO-19, PRTSOLO-20

Recall number
Z-1376-2019
Recalling firm
Terumo Medical Corporation
Product
Portico Solo Re-Collapsible Access System , Model Numbers PRTSOLO-19, PRTSOLO-20
Status
Terminated
Initiated
2019-04-26
Posted
2019-05-28
Terminated
2020-09-22
Reason
There is a potential for dislodgement of the tip from the outer diameter of the sheath resulting in a loss of the smooth transition from the surface of the tip to the outer surface of the expandable sheath.
Root cause
Device Design
Action
Urgent Medical Device Recall letters dated 4/26/19 were sent to customers.
Quantity
1481
Distribution
The products were distributed US nationwide. The products were distributed to the following foreign countries: Australia, Belgium, Brazil, Canada, Portugal, Hong Kong, Peru, Singapore.
Lot, serial or model codes
Lot Codes: VG29 VH28 VK09 VM20 VN08 VP07 WM03 WM31 VG05 VG09 VG21 VK18 VL25 VM11 VN17 VN27 WM03 WM17
510(k) numbers
["K152498"]
PMA numbers
not on file
Product code
DYB
Firm FEI
1118880
Firm city
Elkton
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register