Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer, Inc.: NEXGEN Complete Knee Solution Monoblock Tibial Drill with stop 10.7 mm. Used to prepare cylindrical holes in the native tibia.

Recall number
Z-1376-2013
Recalling firm
Zimmer, Inc.
Product
NEXGEN Complete Knee Solution Monoblock Tibial Drill with stop 10.7 mm. Used to prepare cylindrical holes in the native tibia.
Status
Terminated
Initiated
2013-03-27
Posted
2013-05-24
Terminated
2017-01-26
Reason
Zimmer is initiating a lot specific recall of the Trabecular Metal" Tibia Stop Drill due to the potential of the outside diameter being oversized. As a result, there is a potential for the drill to not pass through the applicable drill guide.
Root cause
Process control
Action
Zimmer sent an "URGENT MEDICAL DEVICE RECALL LETTER" dated April 19, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.
Quantity
69 units
Distribution
Nationwide Distribution.
Lot, serial or model codes
Part No. 00-5887-052-00; lots 62181290 and 62183430
510(k) numbers
["K012866","K020295"]
PMA numbers
not on file
Product code
HTW
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register