Skip to the content

FDA Medical Device Recalls (openFDA)

Animas Corporation: OneTouch(R) Ping(R) Insulin Pump glucose management system. The device is intended to deliver insulin therapy.

Recall number
Z-1376-2012
Recalling firm
Animas Corporation
Product
OneTouch(R) Ping(R) Insulin Pump glucose management system. The device is intended to deliver insulin therapy.
Status
Terminated
Initiated
2012-03-01
Posted
2012-04-06
Terminated
2013-12-24
Reason
There is an error in the pump software where the OneTouch(R) Ping(R) insulin pump did not allow the user to save changes to the time and date on the pump if those changes were made on February 29, 2012 (Leap Day). Users who did not attempt to change the time or date on their pump on February 29, 2012 are not affected by this issue.
Root cause
Software design
Action
The firm initiated their recall on March 1, 2012 and sent Urgent Notification letters dated March 6, 2012 to all consignees by UPS with delivery confirmation. Customers were notified of the issue and the appropriate steps to take in order to mitigate the issue. A Web Letter for users was also posted to the Animas web site. Questions were directed to (855) 230-7577.
Quantity
91,239
Distribution
Worldwide Distribution, (USA) including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of Canada and the territory of Puerto Rico.
Lot, serial or model codes
Part number 101420-00, 10142100, 101422-00, 101424-00, 101425-00, 100430-00, 100431-00, 100432-00, 100434-00, 100435-00, 101420-98, 101421-98, 101422-98, 101424-98, 101425-98, and 10050000.
510(k) numbers
["K080639"]
PMA numbers
not on file
Product code
LZG
Firm FEI
3002920072
Firm city
West Chester
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register