FDA Medical Device Recalls (openFDA)
DePuy Orthopaedics, Inc.: MODULAR CATHCART FRACTURE HEAD HIP BALL, sizes 41mm to 60mm inclusive; metallic femoral head prosthesis, Product No. 136341000, 136342000, 136343000, 136344000, 136345000, 136346000, 136347000, 136348000, 136349000, 136350000, 136351000, 136352000, 136353000, 136354000, 136356000, 136358000, 136360000
- Recall number
- Z-1375-2022
- Recalling firm
- DePuy Orthopaedics, Inc.
- Product
- MODULAR CATHCART FRACTURE HEAD HIP BALL, sizes 41mm to 60mm inclusive; metallic femoral head prosthesis, Product No. 136341000, 136342000, 136343000, 136344000, 136345000, 136346000, 136347000, 136348000, 136349000, 136350000, 136351000, 136352000, 136353000, 136354000, 136356000, 136358000, 136360000
- Status
- Open, Classified
- Initiated
- 2022-05-18
- Posted
- not on file
- Terminated
- not on file
- Reason
- The Cathcart Ball has an inherent +5mm offset which is not currently specified in its Instructions for Use or Surgical Technique Guide. The STG has been updated to reflect the clarification that the +5mm offset should be taken into consideration when using a Cathcart head in a construct for hemi-hip arthroplasty procedures.
- Root cause
- Labeling design
- Action
- On June 13, 2022, the firm notified distributed recall letters to customers titled URGENT FIELD SAFETY NOTICE. Customers were instructed to contact their local sales consultant to receive a copy of the revised STG (reference document number DSUS/JRC/0317/2044 Rev B). Product is not being removed from the field.
- Quantity
- 3488
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Canada, Denmark, Finland, Germany, Greece, Guam, Hong Kong, Ireland, Netherlands, New Zealand, Norway, South Africa, Spain, Sweden, Switzerland, United Kingdom.
- Lot, serial or model codes
- All lots until Instructions for Use (IFU) is revised
- 510(k) numbers
- ["K903084"]
- PMA numbers
- not on file
- Product code
- LZY
- Firm FEI
- 1818910
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28