DePuy Orthopaedics, Inc.: MODULAR CATHCART FRACTURE HEAD HIP BALL, sizes 41mm to 60mm inclusive; metallic femoral head prosthesis, Product No. 136341000, 136342000, 136343000, 136344000, 136345000, 136346000, 136347000, 136348000, 136349000, 136350000, 136351000, 136352000, 136353000, 136354000, 136356000, 136358000, 136360000

Recall number
Z-1375-2022
Recalling firm
DePuy Orthopaedics, Inc.
Product
MODULAR CATHCART FRACTURE HEAD HIP BALL, sizes 41mm to 60mm inclusive; metallic femoral head prosthesis, Product No. 136341000, 136342000, 136343000, 136344000, 136345000, 136346000, 136347000, 136348000, 136349000, 136350000, 136351000, 136352000, 136353000, 136354000, 136356000, 136358000, 136360000
Status
Open, Classified
Initiated
2022-05-18
Posted
not on file
Terminated
not on file
Reason
The Cathcart Ball has an inherent +5mm offset which is not currently specified in its Instructions for Use or Surgical Technique Guide. The STG has been updated to reflect the clarification that the +5mm offset should be taken into consideration when using a Cathcart head in a construct for hemi-hip arthroplasty procedures.
Root cause
Labeling design
Action
On June 13, 2022, the firm notified distributed recall letters to customers titled URGENT FIELD SAFETY NOTICE. Customers were instructed to contact their local sales consultant to receive a copy of the revised STG (reference document number DSUS/JRC/0317/2044 Rev B). Product is not being removed from the field.
Quantity
3488
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Canada, Denmark, Finland, Germany, Greece, Guam, Hong Kong, Ireland, Netherlands, New Zealand, Norway, South Africa, Spain, Sweden, Switzerland, United Kingdom.
Lot, serial or model codes
All lots until Instructions for Use (IFU) is revised
510(k) numbers
["K903084"]
PMA numbers
not on file
Product code
LZY
Firm FEI
1818910
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register