Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc.: Access Immunoassay Systems Assay Protocol File (APF) Software versions for UniCel DxC 600i Systems. Part Number: A54378, UniCel DxC 600i APF, Part Number: A54379, UniCel DxC 600i AAF. The Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentration found in human body fluids.

Recall number
Z-1375-2012
Recalling firm
Beckman Coulter Inc.
Product
Access Immunoassay Systems Assay Protocol File (APF) Software versions for UniCel DxC 600i Systems. Part Number: A54378, UniCel DxC 600i APF, Part Number: A54379, UniCel DxC 600i AAF. The Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentration found in human body fluids.
Status
Terminated
Initiated
2008-09-11
Posted
2012-04-06
Terminated
2012-06-26
Reason
The recall was initiated because Beckman confirmed a customer report that a timing conflict can occur while aspirating samples for the Access Total BhCG assay.
Root cause
Software design
Action
Beckman Coulter sent a "PRODUCT CORRECTIVE ACTION" letter dated September 24, 2008 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers were instructed to share this information with their laboratory staff and retain the notification as part of their laboratory Quality System documentation. Customers were instructed to complete and return the enclosed response form. Customers with questions regarding this notification were instructed to contact Technical Support at (800) 854-3633 or their local Beckman Coulter representative.
Quantity
2948 units total (2762 units in the US)
Distribution
Worldwide Distribution-USA (nationwide) and the country of Canada.
Lot, serial or model codes
Part Number: A54378, UniCel DxC 600i APF,  Lot Numbers: 722631, 823677. Part Number: A54379, UniCel DxC 600i AAF, Lot Numbers: 722630, 823675.
510(k) numbers
["K060256"]
PMA numbers
not on file
Product code
JJE
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register