Varian Medical Systems Oncology Systems: Varis Vision Treatment 6.6/RTP Exchange/Siemens Accelerators

Recall number
Z-1375-06
Recalling firm
Varian Medical Systems Oncology Systems
Product
Varis Vision Treatment 6.6/RTP Exchange/Siemens Accelerators
Status
Terminated
Initiated
2005-09-20
Posted
2006-08-05
Terminated
2006-11-21
Reason
A software anomaly may occur which can lead to patient treatment with the wrong field. The anomaly is reported only to occur when this software version (6.6.5022) is used with Elekta or Siemens linear accelerators .
Root cause
Software design
Action
On 09/20/2005, Varian Medical Systems issued a notification letter via certified mail and FedEx. These letters were issued with instructions on what to do regarding the corrections. Varian Medical Systems intends to release new, updated software.
Quantity
58 units
Distribution
The product was distributed to 34 consignees, located in AZ, CA, MA, MI, NC, PA in the USA, and in Australia, Belgium, Canada, Denmark, Finland, France, Germany, Italy, Lithuania, Spain, Sweden, and the United Kingdom
Lot, serial or model codes
H462074, H462000, H462003, H462004, H462005, H462006, H462011, H462012, H462013, H462014, H462015, H462016, H462017, H462018, H462019, H462020, H462021, H462022, H462023, H462024, H462025, H462026, H462027, H462028, H462029, H462031, H462040, H462045, H462047, H462048, H462049, H462050, H462051, H462052, H462054, H462055, H462057, H462059, H462060, H462061, H462064, H462065, H462066, H462067, H462068, H462072, H462073, H462075, H462076, H462077, H462079, H462082, H462083, H462084, H462085, H462086, H462087
510(k) numbers
["K050479"]
PMA numbers
not on file
Product code
IYE
Firm FEI
2916710
Firm city
Palo Alto
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register