Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems North America Co. Phillips: Philips OmniDiagnost Eleva X-Ray System is a diagnostic imaging device intended for radiographic, fluoroscopic, angiographic, and interventional applications.

Recall number
Z-1374-2008
Recalling firm
Philips Medical Systems North America Co. Phillips
Product
Philips OmniDiagnost Eleva X-Ray System is a diagnostic imaging device intended for radiographic, fluoroscopic, angiographic, and interventional applications.
Status
Terminated
Initiated
2008-03-22
Posted
2008-09-16
Terminated
2010-09-17
Reason
Unexpected movement of the X-ray table may occur. These movements may lead to patient injury, e.g., an uncontrolled tilt may cause the patient to slip off.
Root cause
Software design
Action
On 3/22/08, the firm mailed letters to their customers via certified mail. The letter informed the customers of the recall, advising the customers to, "Use the emergency stop button or release the initiated movement command to halt undesired table movement. If you perform digital and conventional acquisition, you can avoid the problem by closing the single examinations daily or by clearing the patient database regularly." The letter further informed the customers that a Philips representative will be installing updated software at a future date. Contact Sarah Baxter at 425-487-7665 or Philips Call Center at 800-722-9377, #5, #2 if you have questions.
Quantity
22 units
Distribution
Nationwide Distribution -- including states of CA, CO, KY, TN, LA, MI, FL, MS, TX, VA, WI, MO, AZ, and NY.
Lot, serial or model codes
Product number 9896 002 04432. SN number: 43, 49, 52, 107, 123, 131, 152, 157, 171, 179, 185, 202, 266, 333, 337, 389, 408, 446, 485, 491, 514, 515; Site ID #: 506241, 500046, 521648, 522220, 520020, 533112, 533314, 533471, 533530, 535257, 535118, 535924, 540860, 541490, 539106, 547017, 547142, 549798, 547348, 550258, 553676, and 553235.
510(k) numbers
["K032046"]
PMA numbers
not on file
Product code
IZI
Firm FEI
1000524572
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register