Ortho-Clinical Diagnostics: VITROS Performance Verifier II

Recall number
Z-1373-2024
Recalling firm
Ortho-Clinical Diagnostics
Product
VITROS Performance Verifier II
Status
Open, Classified
Initiated
2024-02-27
Posted
2024-03-22
Terminated
not on file
Reason
The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could lead to testing delays and erroneous patient results. This issue caused positive or negative bias in the quantification of sodium (Na+) and total carbon dioxide (ECO2). The incorrect diluent into the Performance Verifier II exhibited a potential negative bias of up to 6 mmol/L for ECO2 and Na+ testing.
Root cause
Under Investigation by firm
Action
Consignees were sent an URGENT PRODUCT CORRECTION NOTIFICATION dated 2/27/24 informing them that for product Lot Q1174, diluent Lot M9815 was found in some sales units instead of the expected diluent Lot Q1175, and for product Lot R1176, diluent Lot P9967 was found in some sales units instead of the expected diluent Lot R1177. Consignees were asked to inspect their inventory to see if any mismatched product from affected lots was present and to discard any affected units. The provided Confirmation of Receipt form is to be returned by 3/8/24. Results received using affected devices are to be reviewed for any unexpected QC shift. The recall notification is to be forwarded if any devices were further distributed and the notification maintained in the User Documentation. Customers with any questions can call Global Services at 1-800-421-3311.
Quantity
11,224 units
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
Lot, serial or model codes
Product Code: 8231474; GTIN: 10758750004577; Lot No. R1176.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JJY
Firm FEI
1000305841
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register