FDA Medical Device Recalls (openFDA)
TriMed Inc.: Hex Driver Part Number HXDRIVER-2.0 AO; Hex Driver Part Number HXDRIVR-2.5 AO; Hex Driver Part Number HXDRIVR-2.5CL AO; Hex Driver Part Number DVHX-2.5/130C AO; Hex Driver Part Number HXDRIVR-3.0 AO; Hex Driver Part Number HXDRIVER-3.5AO
- Recall number
- Z-1373-2018
- Recalling firm
- TriMed Inc.
- Product
- Hex Driver Part Number HXDRIVER-2.0 AO; Hex Driver Part Number HXDRIVR-2.5 AO; Hex Driver Part Number HXDRIVR-2.5CL AO; Hex Driver Part Number DVHX-2.5/130C AO; Hex Driver Part Number HXDRIVR-3.0 AO; Hex Driver Part Number HXDRIVER-3.5AO
- Status
- Terminated
- Initiated
- 2018-01-11
- Posted
- not on file
- Terminated
- 2024-01-18
- Reason
- Following the completion of sterilization cycle(s) of the impacted Hex Driver, the epoxy color band on the impacted Hex Driver can migrate down the shaft and/or potentially onto the sterilization tray or caddy.
- Root cause
- Process control
- Action
- Recall notification letters were sent to affected consignees on January 16, 2018. The consignee letter requests that the consignee return an initial Acknowledgement & Receipt Form to TriMed by January 23, 2018.
- Quantity
- 1525 devices
- Distribution
- US, China, Latin America
- Lot, serial or model codes
- Affected lot numbers: 922361, 922362, 91956, 91957, 92237, 92238, 91858, 92235, 92656, 91856, 91857
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LXH
- Firm FEI
- 3001236812
- Firm city
- Santa Clarita
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28