FDA Medical Device Recalls (openFDA)
Gendex Dental Systems: Orthoralix 8500 DDE Panoramic Dental X-Ray System; Extraoral x-ray unit with timer; Gendex Dental Systems, Des Plaines, IL 60018; part number 110-0234G1 (domestic) and 110-0234G3 (international) The Orthoralix 8500 DDE Panoramic Dental X-Ray System is used as an extraoral source of x-rays for imaging of the dento-maxillofacial area.
- Recall number
- Z-1373-2011
- Recalling firm
- Gendex Dental Systems
- Product
- Orthoralix 8500 DDE Panoramic Dental X-Ray System; Extraoral x-ray unit with timer; Gendex Dental Systems, Des Plaines, IL 60018; part number 110-0234G1 (domestic) and 110-0234G3 (international) The Orthoralix 8500 DDE Panoramic Dental X-Ray System is used as an extraoral source of x-rays for imaging of the dento-maxillofacial area.
- Status
- Terminated
- Initiated
- 2010-12-21
- Posted
- 2011-02-18
- Terminated
- 2011-11-09
- Reason
- The Automatic Exposure Control (AEC) feature located on the control panel of the Orthoralix 8500 Panoramic Dental X-Ray System was functional, but was subsequently disabled in units manufactured starting 5/17/10. North American units manufactured between 12/4/09 and 5/16/10 had functioning AEC buttons, but their manuals stated they were disabled. Units intended for International distribution bet
- Root cause
- Labeling Change Control
- Action
- Gendex sent a recall letter to their accounts on December 21, 2010, stating that the AEC button on the control panel was a non-functional feature that becomes illuminated upon activation and that Gendex would send them a kit to correct this issue and inactivate the button. A second letter was sent to the accounts on January 31, 2011, explaining that a corrective kit was no longer necessary, letting them know if the AEC feature was active or inactive for their system, and providing them with the appropriate user manual addendum for their system. Customers were asked to complete and return the required Acknowledgement Form by fax to 215-997-5665 or e-mail: regulatoryaffairs@gendex.com. For questions regarding this recall call 800-323-8029.
- Quantity
- 574 systems
- Distribution
- Worldwide Distribution - USA including AL, AZ, CA, CO, CT, FL, GA, CO, CT, FL, GA, ID, IL, KS, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NM, NY, OH, OK, PA, TN, TX, VA, WA, and WI and the countries of Australia, Canada, Germany, New Zealand, Taiwan, and Vietnam.
- Lot, serial or model codes
- a) North American units manufactured between 12/4/09 and 5/16/10 and had a working AEC button, but the manual stated that it was disabled: serial numbers 1863844, 1950710, 1950714, 1954576, 1954577, 1954578, 1954579, 1954581, 1954582, 1954583, 1954584, 1954585, 1954587, 1954589, 1954590, 1954591, 1954593, 1954598, 1954599, 1954600, 1954601, 1954603, 1954605, 1954606, 1954607, 1954614, 1954616, 1954617, 1954618, 1954619, 1954622, 1954623, 1954625, 1954626, 1954627, 1954628, 1954629, 1954630, 1954631, 1954632, 1954633, 1954634, 1954635, 1954637, 1954638, 1954639, 1954640, 1954641, 1954642, 1954643, 1954644, 1954645, 1954646, 1954647, 1954648, 1954649, 1954650, 1954651, 1954652, 1954653, 1954654, 1954656, 1954657, 1954658, 1954659, 1954660, 1954661, 1954662, 1954664, 1954665, 1954666, 1954667, 1954668, 1954669, 1954670, 1954673, 1954674, 1954675, 1957474, 1957926, 1957927, 1957929, 1957930, 1957931, 1957932, 1957933, 1957934, 1957935, 1957936, 1957937, 1957940, 1957941, 1957942, 1957943, 1957944, 1957945, 1957946, 1957947, 1957948, 1957949, 1957950, 1957951, 1957952, 1957953, 1957955, 1957956, 1957959, 1957960, 1957963, 1957966, 1957967, 1957968, 1957973, 1957977, 1957978, 1957980, 1957982, 1957983, 1957983, 1957986, 1957987, 1957988, 1957989, 1957991, 1957994, 1957997, 1957998, 1957999, 1958000, 1958003, 1958004, 1958005, 1958006, 1958007, 1958010, 1958011, 1958012, 1958013, 1958014, 1958015, 1958016, 1958017, 1958018, 1958020, 1958021, 1958022, 1958024, 1958025, 1960071, 1960072, 1960073, 1960074, 1961209, 1961210, 1961218, 1961219, 1961226, 1961231, 1961233, 1961234, 1961236, 1961237, 1961243, 1961244, 1961246, 1961247, 1961248, 1961250, 1961253, 1961255, 1961256, 1961257, 1961258, 1961261, 1961263, 1961264, 1961267, 1961268, 1961269, 1961270, 1961271, 1961272, 1961273, 1961274, 1961275, 1961279, 1961280, 1961283, 1961288, 1961289, 1961290, 1961291, 1961293, 1961294, 1961296, 1961298, 1961299, 1961300, 1961306, 1961307, 1963607, 1963608, 1963611, 1963712, 1963713, 1963714, 1963716, 1963717, 1963718, 1963719, 1963720, 1963721, 1963722, 1963725, 1963726, 1963735, 1963737, 1963738, 1963739, 1963740, 1963741, 1963742, 1963743, 1963744, 1963745, 1963746, 1963747, 1963753, 1963756, 1963758, 1963759, 1963760, 1963766, 1963776, 1963810, 1963812, 1963813, 1963814, 1963816, 1963818, 1963819, 1963820, 1963821, 1963822, 1963823, 1963825, 1963826, 1963827, 1963828, 1963830, 1963831, 1963832, 1963833, 1963835, 1963836, 1963837, 1963838, 1963839, 1963840, 1963842, 1963843, 1963848, 1963849, 1963852, 1963853, 1963855, 1963858, 1963860, 1963876, 1963879, 1963883, 1963886, 1963887, 1963888, 1963895, 1963896, 1963898, 1963899, 1963900, 1963902, 1963904, 1963906, 1963909, 1963910, 1964843, 1964845, 1964846, 1970570, 1970571, 1970572, 1970575, 1970576, 1970577, 1970578, 1970580, 1970583, 1970587, 1970606, 1970608, 1970611, 1970612, 1970613, 1970614, 1970615, 1970618, 1970620, 1970621, 1970622, 1970624, 1970625, 1970627, 1970628, 1970629, 1970630, 1970631, 1970632, 1970634, 1970635 and 1970636; b) units manufactured between 5/17/10 and 11/9/10 and had a disabled AEC button, but the English section of the international manual stated that it was functional: serial numbers 1957969, 1957971, 1957974, 1961227, 1961228, 1961260, 1961262, 1961276, 1961278, 1961282, 1961284, 1961285, 1961292, 1961302, 1961303, 1961304, 1961305, 1963723, 1963727, 1963729, 1963750, 1963751, 1963752, 1963754, 1963755, 1963757, 1963762, 1963763, 1963765, 1963767, 1963768, 1963769, 1963774, 1963778, 1963782, 1963783, 1963784, 1963785, 1963788, 1963789, 1963790, 1963791, 1963792, 1963794, 1963795, 1963796, 1963800, 1963803, 1963804, 1963806, 1963808, 1963809, 1963815, 1963861, 1963862, 1963864, 1963870, 1963871, 1963872, 1963877, 1963882, 1963884, 1963894, 1963897, 1963908, 1970568, 1970569, 1970573, 1970574, 1970584, 1970586, 1970589, 1970591, 1970594, 1970595, 1970597, 1970598, 1970599, 1970600, 1970601, 1970603, 1970610, 1970616, 1970619, 1970623 and 1970626; c) units manufactured between 5/17/10 and 11/9/10 and had a disabled AEC button, but the French, German, Italian and Spanish sections of the international manual stated that it was functional: serial numbers 1970638, 1970639, 1970640, 1970641, 1970643, 1970645, 1970649, 1970650, 1970651, 1970652, 1970653, 1970654, 1970663, 1970664, 1973813, 1973814, 1973817, 1973820, 1973824, 1973828, 1973834, 1973848, 1973849, 1973850, 1973851, 1973857, 1973858, 1973878, 1973885, 1973887, 1973889, 1973891, 1973893, 1974792, 1976592, 1976594, 1976598, 1976611, 1976614, 1976619, 1976634, 1976637, 1976638, 1978213, 1978217, 1978221, 1978228, 1978233, 1978252, 1978253, 1978254, 1978255, 1978258, 1978264, 1978265, 1978266, 1978268, 1978269, 1978271, 1978278, 1978279, 1978280, 1978284, 1978285, 1978287, 1978288, 1978289, 1978290, 1978291, 1978292, 1978294, 1978295, 1978296, 1978297, 1978298, 1978299, 1978300, 1978301, 1978302, 1978306, 1978307, 1978308, 1978340, 1978341, 1978342, 1978346, 1978348, 1978350, 1982523, 1982543, 1982552, 1982553, 1982554, 1982561, 1982567, 1982570, 1982571, 1982572, 1982573, 1982575, 1982576, 1982577, 1982578, 1982580, 1982581, 1982582, 1982588, 1982589, 1982590, 1982591, 1982592, 1982593, 1982595, 1982596, 1982597, 1982598, 1982599, 1982601, 1982604, 1982607, 1982608, 1982609, 1982612, 1982614, 1982615, 1982616, 1982617, 1982618, 1982619, 1982621, 1982622, 1982624, 1982625, 1982628, 1982629, 1982631, 1982632, 1982633, 1982634, 1982714, 1982716, 1989268, 1989272, 1989275, 1989276, 1989278, 1989281, 1989293, 1989294, 1989295, 1989298, 1989303, 1989306, 1989307, 1989314, 1989315, 1989316, 1989328, 1989331, 1989339, 1989340, 1989389, 1989395, 1989396, 1989400, 1989401, 1989403, 1989406 and 1989407
- 510(k) numbers
- ["K050255"]
- PMA numbers
- not on file
- Product code
- EHD
- Firm FEI
- 3004115000
- Firm city
- Des Plaines
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28