Biotronik Inc: Lumax Cardiac Resyncronization Therapy Defibrillator (CRT-D); Implantable, dual-chamber defibrillator; Model Number: HF-T; Catalog Number 355263; Software Version 603.U; Catalog Model Number 359251; ICS 3000 units are composed of three Catalog/Model Numbers: 349527, 349530, and 355547. The model 349530 is the component which is loaded/updated with software.

Recall number
Z-1373-2008
Recalling firm
Biotronik Inc
Product
Lumax Cardiac Resyncronization Therapy Defibrillator (CRT-D); Implantable, dual-chamber defibrillator; Model Number: HF-T; Catalog Number 355263; Software Version 603.U; Catalog Model Number 359251; ICS 3000 units are composed of three Catalog/Model Numbers: 349527, 349530, and 355547. The model 349530 is the component which is loaded/updated with software.
Status
Terminated
Initiated
2007-01-10
Posted
2008-07-29
Terminated
2011-05-05
Reason
Timing Feature Error: An implementation error allows the usage of the V-V timing feature, which is not approved for use in the U.S.
Root cause
Software design
Action
Consignees were notified by a letter sent 1/10/07. The letter advised users to stop usage of the affected software. Users were instructed not to implant additional devices until the software program has been corrected. The recalling firm requested that all affected software be scrapped or returned to them. Users were requested update the software with a corrected version as soon as it becomes available.
Quantity
53 of the 603.U discs distributed.
Distribution
Nationwide Distribution including states of AL, AZ, CA, CO, DE, FL, GA, IL, LA, MI, MO, NC, NJ, NV, NY, OH, OR, PA, SC, TX and WA.
Lot, serial or model codes
Software version 603.U, Catalog/Model Number: 359251; ICS 300 programmers Catalog/Model Numbers: 349527, 449530 and 355547; Lumax 340 HF-T Units, Catalog Number: 355263
510(k) numbers
not on file
PMA numbers
["P050023S001","P050023S006"]
Product code
MRM
Firm FEI
1028232
Firm city
Lake Oswego
Firm state
OR
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register