FDA Medical Device Recalls (openFDA)
Hologic, Inc.: Aptima CMV Quant Assay, Reference numbers: AW-22600-001, AW-25509-001, and AW-26381-001 Amplification Reagent, Enzyme Reagent, Promoter Reagent, Amplification Reconstitution Solution, Enzyme Reconstitution Solution, Promoter Reconstruction solution, Target capture Reagent, Reconstitution Collars.
- Recall number
- Z-1371-2024
- Recalling firm
- Hologic, Inc.
- Product
- Aptima CMV Quant Assay, Reference numbers: AW-22600-001, AW-25509-001, and AW-26381-001 Amplification Reagent, Enzyme Reagent, Promoter Reagent, Amplification Reconstitution Solution, Enzyme Reconstitution Solution, Promoter Reconstruction solution, Target capture Reagent, Reconstitution Collars.
- Status
- Open, Classified
- Initiated
- 2023-09-14
- Posted
- 2024-03-22
- Terminated
- not on file
- Reason
- Use of assay along with plasma specimen may result in error code which would invalidate sample.
- Root cause
- Use error
- Action
- On February 23, 2024, Hologic issued a "Urgent Medical Device Recall Update" via FedEx. The update notified consignees of a sample dilution workflow for specific plasma specimens. On September 14, 2023 Hologic issued a "Urgent Medical Device Recall Notification" to affected consignees, via Fed Ex. Hologic provided the following information to customers: 1. Continue to use the Aptima CMV Quant assay. 2.When obtaining an ML2 error, please run the Mag Wash Clean Procedure and Contact Hologic Technical Support 3. With the Aptima CMV Quant assay, it is not recommended to retest a sample that was previously invalidated due to an ML2 error on the Panther instrument. 4. Complete Customer Response Form On (date) the firm issued a Customer Technical Bulletin, to consignees describing a sample dilution workflow for plasma specimens.
- Quantity
- 684 kits
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, CT, FL, ID, IL, MD, MI, MO, TN, TX, WI and the countries of Australia, Germany, Denmark, Spain, France, Great Britain, Italy, Norway, Qatar, Sweden.
- Lot, serial or model codes
- Lot #: 402941/ UDI:15420045514553
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- PAB
- Firm FEI
- 2024800
- Firm city
- San Diego
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28