Skip to the content

FDA Medical Device Recalls (openFDA)

Hamilton Medical AG: Hamilton-C2 Ventilator Unit Intended to provide ventilator support to adults and pediatrics and optionally infants and neonates in the hospital.

Recall number
Z-1371-2018
Recalling firm
Hamilton Medical AG
Product
Hamilton-C2 Ventilator Unit Intended to provide ventilator support to adults and pediatrics and optionally infants and neonates in the hospital.
Status
Open, Classified
Initiated
2018-03-12
Posted
not on file
Terminated
not on file
Reason
After two and a half years, the battery fuel gauge may indicate a higher battery charge than what is available.
Root cause
Other
Action
Customers were notified via letter on about 03/12/2018. The letter identified the affected device and described the issue. Instructions for medical facilities are to inspect the batteries for the battery manufacturing date and replace any batteries that are found to be in use for two years or more. Instructions for determining the battery manufacturing date are included in the Medical Device Field Safety Corrective Action Update attached to the letter. Questions or concerns can be directed to Bret Everett at 817-909-0308 or Hamilton Technical Support at 1-800-426-6331.
Quantity
739 units total
Distribution
Worldwide Distribution -- USA, including the states of AZ, CA, CO, FL, GA, HI, IL, IN, KY, MA, MI, MN, MO, MT, NC, ND, NE, NV, OH, OK, PA, SC, SD, TN, TX, UT, VA, VT, and WA; and, the countries of Algeria, Angola, Argentina, Armenia, Australia, Austria, Austria, Azerbaijan, Bangladesh, Belarus, Belgian, Bolivia, Brazil, Bulgaria, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czechia, Denmark, Dominican. Rep., Ecuador, Egypt, El Salvador, Estland, Finland, France, Gabon, Germany, Greece, Guatemala, Holland, Honduras, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Mauritius, Macedonia, Mexico, Mongolia, Morocco, Myanmar, Nepal, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, Rwanda, Saudi Arabia, Scotland, Serbia, Seychelles, Slovakia, Slovenia, South Africa, Spain, SRI Lanka, Sudan, Suriname, Sweden, Sweden, Switzerland, Syria, Taiwan, Tajikistan, Thailand, Trinidad & Tobago, Tunisia, Turkey, UAE, Uganda, UK , Ukraine, Uruguay, Venezuela, Vietnam, and Yemen.
Lot, serial or model codes
Product No. 160001  All Serial Numbers.
510(k) numbers
["K121225"]
PMA numbers
not on file
Product code
CBK
Firm FEI
3001421318
Firm city
Bonaduz
Firm state
not on file
Firm country
Switzerland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register