FDA Medical Device Recalls (openFDA)
Edwards Lifesciences, LLC: Edwards Lifesciences Fem-Flex II Femoral Arterial Cannula 8, 10, 12 French,Sterile EO, Rx Only, Manufacturer Edwards Lifesciences LLC, Irvine, CA.-Model Numbers: FEMII008A, FEMII008AT, FEMII008V, FEMII010A, FEMII010AT, FEMII010V, FEMII012A, FEMII012AT, and FEMII012V. Edwards Femoral Access Cannulae are intended to provide a means of draining the blood flow (venous), or perfusing blood into the body (arterial) of a patient during cardiopulmonary bypass procedures.
- Recall number
- Z-1371-2015
- Recalling firm
- Edwards Lifesciences, LLC
- Product
- Edwards Lifesciences Fem-Flex II Femoral Arterial Cannula 8, 10, 12 French,Sterile EO, Rx Only, Manufacturer Edwards Lifesciences LLC, Irvine, CA.-Model Numbers: FEMII008A, FEMII008AT, FEMII008V, FEMII010A, FEMII010AT, FEMII010V, FEMII012A, FEMII012AT, and FEMII012V. Edwards Femoral Access Cannulae are intended to provide a means of draining the blood flow (venous), or perfusing blood into the body (arterial) of a patient during cardiopulmonary bypass procedures.
- Status
- Terminated
- Initiated
- 2015-03-23
- Posted
- 2015-04-02
- Terminated
- 2017-04-17
- Reason
- Edwards Lifesciences is recalling Fem-Flex II Pediatric Femoral Arterial Cannula sizes 8, 10, 12 French because of the potential of tissue damage caused by a protruding wire located at the tip of the cannula.
- Root cause
- Process change control
- Action
- The firm, Edwards, sent an "URGENT - PRODUCT RECALL - ACTION REQUIRED" letter dated March 23, 2015 via FED EX on March 27, 2015 to their customers. The letter described the product, problem, and actions to be taken. The customers were instructed to review entire inventory for the lots listed; complete and return attached acknowledgment form via fax to Edwards Customer Service at 800.422.9329. within three days of receipt of this Field Safety Notice; contact Customer Service at 800.422.3278 to obtain an RGA number and replacement product; return affected product to Edwards Lifesciences, Attn: Cirilo Chaparro, 12050 Lone Peak Drive, Draper, UT 84020, Attention: RECALL, RGA#XXX; and transfer this notice to other organizations if the affected devices have been transferred to any another facilities. If you have any questions that have not been answered by this letter, please call Edwards Customer Services at 800.424.3278 from the hours of 6:00AM - 4:30PM PST; Edwards Customer Service at (800) 268-3993 from 8:00AM 4:30PM Eastern Time or contact your Edwards sales representative concerning the recall.
- Quantity
- 5125 units
- Distribution
- Nationwide Distribution.
- Lot, serial or model codes
- Model/Lot Numbers: FEMII008A/59751073, 59775775, 59775776, 59775777 & 59852930; FEMII008AT/59807985, 59867050, & 59873263; FEMII008V/59873250 & 59873251; FEMII010A/59740468, 59773806, 59792415, 59792416 & 59852934; FEMII010AT/ 59747819, 59807986,59852935, 59890916 & 59896910; FEMII010V/59751074, 59849119 & 59890924; FEMII012A/ 59801792,59867064, 59884766 & 59884778 FEMII012AT/59852940 & 59867051; and FEMII012V/59723307, 59796683, 59849124 & 59873252
- 510(k) numbers
- ["K140208"]
- PMA numbers
- not on file
- Product code
- DWF
- Firm FEI
- 1713910
- Firm city
- Draper
- Firm state
- UT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28