Skip to the content

FDA Medical Device Recalls (openFDA)

Baxa Corporation: Exacta-Mix 600 Pharmacy Compounding System is used in the pharmacy to compound multiple source ingredients into on final solution based on user-specified sequences and volumes.

Recall number
Z-1371-2008
Recalling firm
Baxa Corporation
Product
Exacta-Mix 600 Pharmacy Compounding System is used in the pharmacy to compound multiple source ingredients into on final solution based on user-specified sequences and volumes.
Status
Terminated
Initiated
2008-02-11
Posted
2008-08-19
Terminated
2012-03-16
Reason
Flush solution from TPN compounder may be added to patient's TPN bag.
Root cause
Software design
Action
Consignees were notified by letter (Exacta-Mix Compounder Software Safety Alert) on 02/11/2008 advising them of the necessity to install a waste bag prior to a machine flush.
Quantity
221 units
Distribution
Nationwide, including military facilities in CA, FL, HI, MS, NC, TX, VA and VA facilities in NY, OR, SC, UT. Foreign distribution to Australia, Canada, China and New Zealand.
Lot, serial or model codes
Model 071B-RS232, all serial numbers
510(k) numbers
not on file
PMA numbers
not on file
Product code
NEP
Firm FEI
1419106
Firm city
Englewood
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register