FDA Medical Device Recalls (openFDA)
Alcon Research, LLC: SYRINGE,BULB,2OZ/60CC,GREEN, Part: 300055070, packaged inside Alcon Custom-Pak Surgical Procedure Packs, Products: 18018-06, 3865-44, 5675-76, 13554-11, 18018-06, 15190-16 , 15190-16, 15190-16, 15190-16, 15190-16, 13554-11, 5918-39, 15190-16, 15190-16, 9208-66, 9208-66, 5675-76, 5918-39, 9208-66, 15190-16, 9065-25, 5918-39, 16066-15, 16066-15
- Recall number
- Z-1370-2024
- Recalling firm
- Alcon Research, LLC
- Product
- SYRINGE,BULB,2OZ/60CC,GREEN, Part: 300055070, packaged inside Alcon Custom-Pak Surgical Procedure Packs, Products: 18018-06, 3865-44, 5675-76, 13554-11, 18018-06, 15190-16 , 15190-16, 15190-16, 15190-16, 15190-16, 13554-11, 5918-39, 15190-16, 15190-16, 9208-66, 9208-66, 5675-76, 5918-39, 9208-66, 15190-16, 9065-25, 5918-39, 16066-15, 16066-15
- Status
- Open, Classified
- Initiated
- 2024-02-21
- Posted
- 2024-03-22
- Terminated
- not on file
- Reason
- Medline Bulb irrigation syringes recalled due to potential for packaging sterile barrier to be breached, and these syringes were packed inside surgical procedure packs that were subsequently terminally sterilized, so after the packs are opened, syringes should be discarded at surgical case end, and not reshelved, because the use of non-sterile bulb irrigation syringes may increase infection risk.
- Root cause
- Nonconforming Material/Component
- Action
- On 2/21/24, correction notices were mailed to customers who were asked to do the following: 1) Review your inventory to determine if you have any affected product within your facility. 2) If you have any affected Custom Pak lots in inventory, please use the included stickers to label any affected product. If you have previously reshelved either of the affected Medline Bulb Irrigation Syringe lots, dispose them. 3) All components inside the Pak are intended to be used only for one specific surgical case and should be discarded at the end of the surgical case. 4) Forward the correction notice to all departments within your organization who may be in possession of affected product; and any other organization to which this product may have been transferred. 5) Complete and return the response form via email to Market.Actions@Alcon.com Customers with questions or concerns can call the firm at 1-800-862-5266.
- Quantity
- 886
- Distribution
- US Nationwide distribution in the states of OH, WV.
- Lot, serial or model codes
- Product/Lot/Expiration/UDI-DI: 18018-06/15T707/02/28/2025/380655148440, 3865-44/15T2RC/02/28/2025/380655148150, 5675-76/15T70E/02/28/2025/380655148174, 13554-11/15V4RC/02/28/2025/380655143711, 18018-06/15V3E0/02/28/2025/380655148440, 15190-16/JZ5179517/06/30/2024/380655148303, 15190-16/JZ5179519/06/30/2024/380655148303, 15190-16/JZ5179521/06/30/2024/380655148303, 15190-16/JZ5179523/06/30/2024/380655148303, 15190-16/JZ5179529/06/30/2024/380655148303, 13554-11/15TAUD/01/31/2026/380655143711, 5918-39/15TE5F/02/28/2025/380655149140, 15190-16/15THVA/04/30/2025/380655148303, 15190-16/15TNXJ/04/30/2025/380655148303, 9208-66/15V8CU/04/30/2025/380651693326, 9208-66/15XVDD/05/31/2025/380651693326, 5675-76/JZ5180811/06/30/2024/380655148174, 5918-39/JZ5180812/06/30/2024/380655149140, 9208-66/JZ5180332/03/31/2025/380651693326, 15190-16/JZ5180856/05/31/2025/380655148303, 9065-25/JZ5187593/11/30/2025/380651682849, 5918-39/JZ5187260/06/30/2024/380655149140, 16066-15/JZ5190913/07/31/2025/380655149683, 16066-15/JZ5191891/07/31/2025/380655149683 Bulb Syringe lots: 96922110001 or 96923100001
- 510(k) numbers
- ["K880961"]
- PMA numbers
- not on file
- Product code
- LRO
- Firm FEI
- 1000117501
- Firm city
- Houston
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as