Skip to the content

FDA Medical Device Recalls (openFDA)

Cook Inc.: Cook Salle Intraoperative Pyeloplasty Stent Set PRODUCT/CATALOG NUMBER/GPN: 4.0 Fr - SIPSF-040018-56-6 G18168 4.0 Fr - SIPSF-040018-59 G32773 4.7 Fr SIPSF-050018-59 G32774

Recall number
Z-1370-2019
Recalling firm
Cook Inc.
Product
Cook Salle Intraoperative Pyeloplasty Stent Set PRODUCT/CATALOG NUMBER/GPN: 4.0 Fr - SIPSF-040018-56-6 G18168 4.0 Fr - SIPSF-040018-59 G32773 4.7 Fr SIPSF-050018-59 G32774
Status
Terminated
Initiated
2019-04-22
Posted
not on file
Terminated
2020-05-05
Reason
Kwart RetroInject Stent Set (Multi-Length) and Salle Intraoperative Pyeloplasty Stent Set Instructions for Use (IFU) did not include a warning for stent knotting.
Root cause
Error in labeling
Action
Cook issued MEDICAL DEVICE CORRECTION letters via courier to US Consignees on April 22, 2019. The letter provides reason for recall, products affected, health risk, action to take: Understand that stent knotting is a potential complication associated with use of the Kwart Retro-Inject" Stent Set and Salle Intraoperative Pyeloplasty Stent Set and should be considered if significant resistance is encountered during attempts at removal. 2. Please maintain a copy of this notice with the current IFU or product(s) in your inventory. 3. If you have affected product in inventory, you may continue to use these products. Cook Medical is not requesting product be returned. 4. Complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter.5. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected products have been transferred. Questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235
Quantity
3711
Distribution
Nationwide Foreign: Australia Canada Austria Belgium Denmark Germany Spain Finland France Martinique French Polynesia Switzerland Ireland Italy Netherlands Poland Sweden United Kingdom Hong Kong New Zealand Chile Colombia Israel Russian Federation Czech Republic Greece Jordan Slovakia South Africa United Arab Emirates Slovenia Lithuania Morocco Romania Qatar Bulgaria Libya Saudi Arabia Turkey Ukraine Thailand
Lot, serial or model codes
All lot codes
510(k) numbers
["K810368"]
PMA numbers
not on file
Product code
KOB
Firm FEI
1820334
Firm city
Bloomington
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register