Skip to the content

FDA Medical Device Recalls (openFDA)

BD Biosciences, Systems & Reagents: BD FACSCalibur; Catalog Numbers 342973, 342975, 34976 In vitro Diagnostics Use.

Recall number
Z-1370-2015
Recalling firm
BD Biosciences, Systems & Reagents
Product
BD FACSCalibur; Catalog Numbers 342973, 342975, 34976 In vitro Diagnostics Use.
Status
Terminated
Initiated
2015-03-06
Posted
2015-06-26
Terminated
2016-04-04
Reason
The circuit breakers on some BD FACSCalibur instruments may have the incorrect rated amperage (20 amps instead of 2 amps) causing the circuit breaker to fail to open in an over-current situation.
Root cause
Mixed-up of materials/components
Action
BD Biosciences sent an Important Product Recall letter on March 6, 2015 by certified mail and Field Change Notification (FCN FCB15-01-660312 directing Field Service Engineers to correct the circuit breaker is in process. Customers with questions were instruct to contact BD Customer Support at 1-877-232-8995 (prompt 2) in the United States. For customers outside the US were instructed to contact their local BD Biosciences representative or distributor. For questions regarding this recall call 408-954-6080.
Quantity
72 devices
Distribution
Worldwide Distribution -US including TX, NJ, and OK and Internationally to Hong Kong, Japan, Singopore, and Belguim
Lot, serial or model codes
Serial Nos. E34297300656, E34297300657, E34297300659, E34297300660, E34297300661, E34297300666, E34297300667, E34297300668, E34297300669, E34297300670, E34297300671, E34297300673, E34297300675, E34297300676, E34297300677, E34297300678  E34297502563, E34297502564, E34297502565, E34297502566, E34297502567, E34297502568, E34297502569, E34297502570, E34297502571, E34297502572, E34297502573, E34297502574, E34297502575, E34297502576, E34297502577, E34297502578, E34297502579, E34297502580, E34297502581, E34297502583, E34297502584, E34297502585, E34297502586, E34297502587, E34297502588, E34297502589, E34297502590, E34297502591, E34297502592, E34297502593, E34297502594, E34297502595, E34297502596, E34297502597, E34297502599, E34297502600, E34297502601, E34297502602, E34297502603, E34297502604, E34297502605, E34297502606, E34297502607, E34297502608, E34297502612, E34297502613, E34297502619, E34297502620, E34297502621, E34297502629, E34297600499, E34297600501, E34297600502, E34297600503, E34297600507, E34297600508.
510(k) numbers
["K980858"]
PMA numbers
not on file
Product code
OYE
Firm FEI
2916837
Firm city
San Jose
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register