Skip to the content

FDA Medical Device Recalls (openFDA)

Hologic, Inc.: Hologics SecurView DX workstation, software version 7.3.0 intended for viewing mammography images and other imaging modalities. SecurView DX is intended for selection, display, manipulation, filing and media interchange of multimodality images from a variety of different modality systems. It also interfaces to various image storage and printing devices, using DICOM or similar interface standards. The device used with FDA-cleared monitors may be used by a trained physician for display, manipulation, and interpretation of lossless compressed or non-compressed mammographic images for screening and diagnostic mammography, as well as any other DICOM multimodality image.

Recall number
Z-1370-2012
Recalling firm
Hologic, Inc.
Product
Hologics SecurView DX workstation, software version 7.3.0 intended for viewing mammography images and other imaging modalities. SecurView DX is intended for selection, display, manipulation, filing and media interchange of multimodality images from a variety of different modality systems. It also interfaces to various image storage and printing devices, using DICOM or similar interface standards. The device used with FDA-cleared monitors may be used by a trained physician for display, manipulation, and interpretation of lossless compressed or non-compressed mammographic images for screening and diagnostic mammography, as well as any other DICOM multimodality image.
Status
Terminated
Initiated
2012-02-28
Posted
2012-04-05
Terminated
2012-05-07
Reason
SecurView DX with software version 7.3.0 defects when used to view mammography tomosynthesis images
Root cause
Software design
Action
Hologic Inc. sent an "URGENT: MEDICAL DEVICE RECALL" letter dated February 28, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter provides instructions on how to re-review the stored images and correct any annotation defect. A SecurView DX 7.3.0 Confirmation of Notification form was enclosed for customers to complete and return to the firm via fax at 866-652-8674. Contact Hologic's Help Desk at 1-877-371-4372 for questions regarding this notice.
Quantity
7 units
Distribution
Worldwide Distribution-Distributed in the state of Texas and the countries Belgium, Brazil, and Canada.
Lot, serial or model codes
software version 7.3.0 S/N:298051183669, 29703060119, 298031083133, 298120882443, 298090882384, 29803070916, 2981070859
510(k) numbers
["K103385"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
1221300
Firm city
Bedford
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register