Roche Diagnostics Corp.: ACCU-CHEK Go Care Kit; Catalog number 04470052001. Kit includes an ACCU-CHEK Go meter.

Recall number
Z-1370-05
Recalling firm
Roche Diagnostics Corp.
Product
ACCU-CHEK Go Care Kit; Catalog number 04470052001. Kit includes an ACCU-CHEK Go meter.
Status
Terminated
Initiated
2005-06-07
Posted
2005-08-13
Terminated
2007-07-02
Reason
The meter may self-start without a sample having been applied and report erroneous results without an error message if an undosed test strip is bent upwards, and held upwards, while the meter is waiting for the sample.
Root cause
Other
Action
Recall letters dated 6/7/05 were sent to all customers. Distributors were requested to send a copy of the notification to all customers of test strips used with the product. Users are instructed to never operate the meter without the optics cover in place, and this information will be included in test strip boxes.
Quantity
2490
Distribution
Nationwide.
Lot, serial or model codes
All units.
510(k) numbers
["K040796"]
PMA numbers
not on file
Product code
NBW
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register