Roche Diagnostics Corp.: ACCU-CHEK Go Care Kit; Catalog number 04470052001. Kit includes an ACCU-CHEK Go meter.
- Recall number
- Z-1370-05
- Recalling firm
- Roche Diagnostics Corp.
- Product
- ACCU-CHEK Go Care Kit; Catalog number 04470052001. Kit includes an ACCU-CHEK Go meter.
- Status
- Terminated
- Initiated
- 2005-06-07
- Posted
- 2005-08-13
- Terminated
- 2007-07-02
- Reason
- The meter may self-start without a sample having been applied and report erroneous results without an error message if an undosed test strip is bent upwards, and held upwards, while the meter is waiting for the sample.
- Root cause
- Other
- Action
- Recall letters dated 6/7/05 were sent to all customers. Distributors were requested to send a copy of the notification to all customers of test strips used with the product. Users are instructed to never operate the meter without the optics cover in place, and this information will be included in test strip boxes.
- Quantity
- 2490
- Distribution
- Nationwide.
- Lot, serial or model codes
- All units.
- 510(k) numbers
- ["K040796"]
- PMA numbers
- not on file
- Product code
- NBW
- Firm FEI
- 1823260
- Firm city
- Indianapolis
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28