Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corporation: Brand Name: Spectrum IQ Product Name: Infusion Pump Model/Catalog Number: 3570009 Software Version: N/A Product Description: Infusion Pump, one unit per carton Component: N/A

Recall number
Z-1368-2026
Recalling firm
Baxter Healthcare Corporation
Product
Brand Name: Spectrum IQ Product Name: Infusion Pump Model/Catalog Number: 3570009 Software Version: N/A Product Description: Infusion Pump, one unit per carton Component: N/A
Status
Open, Classified
Initiated
2026-01-13
Posted
2026-02-13
Terminated
not on file
Reason
Pumps were released without full testing being performed, including occlusion alarm testing.
Root cause
Employee error
Action
Firm notified the single customer on 1/13/26 via phone call. Customer was asked to work with on-site service technician to coordinate return of affected product.
Quantity
5
Distribution
States: Ohio Countries: United States
Lot, serial or model codes
Lot Code: GTIN 00085412610900, Serial Numbers 3815663, 3823468, 3824743, 3825234, and 3825599
510(k) numbers
["K251636"]
PMA numbers
not on file
Product code
FRN
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register