Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems North America Co. Phillips: Allura XPER FD 10 and Field Extension for XPER Cardiovascular System and Allura XPER FD 20 and Allura CV20. Angiographic X-ray systems with a solid state X-ray imaging device for cardiovascular and vascular diagnostic and interventional procedures.

Recall number
Z-1368-2011
Recalling firm
Philips Medical Systems North America Co. Phillips
Product
Allura XPER FD 10 and Field Extension for XPER Cardiovascular System and Allura XPER FD 20 and Allura CV20. Angiographic X-ray systems with a solid state X-ray imaging device for cardiovascular and vascular diagnostic and interventional procedures.
Status
Terminated
Initiated
2010-11-04
Posted
2011-03-22
Terminated
2011-07-21
Reason
Adjustment of the micro-switches in the footswitch used on the Philips cardiovascular X-ray systems was not implemented. If it is programmed for exposure, it is possible that after pressing the pedal on the footswitch it does not release itself. The patient may be exposed to X-ray longer than intended.
Root cause
Pending
Action
On 11/5/10, Philips began sending the Electronic Product Radiation Warning notification letter, dated November 4, 2010, to their consignees. The consignees were informed that the middle pedal of the foot switch of the Allura FD10, FD20, CV20 systems, and the Allura Field Extension series does not not always become inactive (sticks) when pedal is released. The function of the middle pedal on the foot switch is freely programmable by the customer. If it is programmed for exposure and if the pedal does not release by itself, the patient will be exposed to X-ray longer than intended. The consignees were advised to check if their footswitch is possibly affected. Footswitch identification 452270000141 is noted. If the pedal does not release by itself press the red emergency stop button on the system. Consignees can call Philips Healthcare Call Center at 800-722-9377, #5 and Reference FCO 72200189.
Quantity
4 units
Distribution
Worldwide Distribution -- US, including states of CA, IL, MA, NC, and OC.
Lot, serial or model codes
Site Numbers: 104075, 104497, 537745, and 51333258.
510(k) numbers
["K033737","K041949"]
PMA numbers
not on file
Product code
IZI
Firm FEI
1000524572
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register