Skip to the content

FDA Medical Device Recalls (openFDA)

Gambro Renal Products, Inc.: Gambro cartridge blood tubing set used with Gambro Dasco Phoenix and Cobe Centrysystem 3 (and 3+) Hemodialysis System, Gambro, Lakewood, CO 80401

Recall number
Z-1367-2008
Recalling firm
Gambro Renal Products, Inc.
Product
Gambro cartridge blood tubing set used with Gambro Dasco Phoenix and Cobe Centrysystem 3 (and 3+) Hemodialysis System, Gambro, Lakewood, CO 80401
Status
Terminated
Initiated
2007-09-10
Posted
2008-03-25
Terminated
2012-01-03
Reason
Kinked tubing on hemodialysis device may cause hemolysis.
Root cause
Device Design
Action
On September 10, 2007, the firm sent an Advisory Notice to all consignees alerting them that a kink in the tubing can result in physical hemolysis. The firm included with the Notice a Quick Reference Guide that provides drawing of the proper installation of the Phoenix and describes how to prevent bloodline kinking. A customer reply form was also included. The firm is in the process of making labeling changes that will be released with the updated software revisions.
Quantity
41,000,000
Distribution
Nationwide, including 17 VA facilities and 1 military facility. Foreign distribution to Canada, Korea, Mexico.
Lot, serial or model codes
All catalog numbers involving the Gambro Cartridge Blood Sets: 003409-400, 003409-410, 003410-500, 003410-510, 004310-710, 003411-500, 003412-500, 003414-500, 003414-510, 003415-510, 003422-500, 003422-515, 003422-520, 003429-500.
510(k) numbers
["K001156"]
PMA numbers
not on file
Product code
KDI
Firm FEI
1713683
Firm city
Lakewood
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register