Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Molecular Systems Inc: Amplicor [3] AV-HRP, Avidin-HRP (Horseradish Peroxidase) BGG Conjugate. Kit: AMPLICOR MTB Detection Kit, Gen 2 Export IVD/CE.

Recall number
Z-1367-04
Recalling firm
Roche Molecular Systems Inc
Product
Amplicor [3] AV-HRP, Avidin-HRP (Horseradish Peroxidase) BGG Conjugate. Kit: AMPLICOR MTB Detection Kit, Gen 2 Export IVD/CE.
Status
Terminated
Initiated
2004-07-07
Posted
2004-08-19
Terminated
2005-01-07
Reason
An increased frequency of 'blue foci' that potentially can cause elevated A450 background in microwell plate wells after the addition of conjugate reagent during PCR detection.
Root cause
Other
Action
With respect to U.S distribution, email/letter on 6/24/04 to the local safety officers and general managers stationed at the affiliated distribution center, Roche Diagnostics Corporation of Indianapolis, IN. With respect to overseas distributors, email/letter on 6/24/04 to the local safety officers and general managers stationed at the affiliated distribution centers. Roche stated that the local safety officer is responsible to communicate the advisory both to the customers/end-users and the regulatory authority of the country. On July 6, 2004, a Revised Important Product Advice Notice was sent out to affiliates requiring the immediate discontinuation of the use of the affected kits, and destruction of any kits remaining in customers hands.
Quantity
189 kits
Distribution
♦ Amplicor CT/NG CT Detection Kit, Lot No.E12231, Material No.20759392018, Expiration 2/2005 , Shipped To: Roche Diagnostics Corporation, 9115 Hague Road, P.O. Box 50457 Indianapolis, IN 46250-0457. ♦ Amplicor CT/NG NG Detection Kit, Lot No.E12647, Material No.20759406018, Expiration 2/2005, Shipped To: Roche Diagnostics Corporation, 9115 Hague Road, P.O. Box 50457 Indianapolis, IN 46250-0457. ♦ Amplicor HCV Detection Kit, v2.0, Lot No.F01023, Material No.21118439018, Expiration 5/2005, Shipped To: Roche Diagnostics Corporation, 9115 Hague Road, P.O. Box 50457 Indianapolis, IN 46250-0457. ♦ Amplicor IC Detection Kit, 96 tests, Lot No.E09554, Material No.20751952018, Expiration 5/2005, Shipped To: Roche Diagnostics Corporation, 9115 Hague Road, P.O. Box 50457 Indianapolis, IN 46250-0457. Export-U.S Distributed Products: AmpliCap HCV Monitor Test, Lot No.E09193, Material No.21111337018, Expiration 8/2004 , Shipped To: Roche Diagnostics KK, Nippon Roche KK, Kamakura Factory 200, Aza Sotokochi, Kajiwara, Kamakura City, Kanagawa Pref, Japan. Amplicor IC Detection Kit, Lot No.E13115, Material No.20763306122, Expiration 7/2005, Shipped To: Roche Diagnostics KK, Nippon Roche KK, Kamakura Factory 200, Aza Sotokochi, Kajiwara, Kamakura City, Kanagawa Pref, Japan; Shipped To: Roche Diagnostics Canada, 201 Boulevard Armand-Frappier, Laval QC H7V 4A2, Canada; Shipped To: Roche Diagnostics GmbH, GX-OSW, Sandhoferstr. 116, D-68305 Mannheim. Amplicor M. avium Detection Kit, Lot No.E12990, Material No.20759880018, Expiration 11/2004, Shipped To: Roche Diagnostics KK, Nippon Roche KK, Kamakura Factory 200, Aza Sotokochi, Kajiwara, Kamakura City, Kanagawa Pref, Japan. Amplicor MTB Detection Kit, Lot No.E13307, Material No.20757462122, Expiration 9/2004, Shipped To: Roche Diagnostics KK, Nippon Roche KK, Kamakura Factory 200, Aza Sotokochi, Kajiwara, Kamakura City, Kanagawa Pref, Japan; Shipped To: Roche Diagnostics Canada, 201 Boulevard Armand-Frappier, Laval QC H7V 4A2, Canada Shipped To: Roche Diagnostics GmbH, GX-OSW, Sandhoferstr. 116, D-68305 Mannheim. Amplicor NG Detection Kit, Lot No.E10894, Material No.20744026122, Expiration 2/2005, Shipped To: Roche Diagnostics KK, Nippon Roche KK, Kamakura Factory 200, Aza Sotokochi, Kajiwara, Kamakura City, Kanagawa Pref, Japan ; Shipped To: Roche Diagnostics Canada, 201 Boulevard Armand-Frappier, Laval QC H7V 4A2, Canada; Shipped To: Roche Diagnostics GmbH, GX-OSW, Sandhoferstr. 116, D-68305 Mannheim. There are no Govt. accounts.
Lot, serial or model codes
Amplicor [3] AV-HRP, Lot No. E09569 found in kit lot:  Lot E13307, Material No. 20757462122, Exp. 9/2004.
510(k) numbers
not on file
PMA numbers
not on file
Product code
MWA
Firm FEI
1000121742
Firm city
Belleville
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register