Skip to the content

FDA Medical Device Recalls (openFDA)

Draeger Medical Systems, Inc.: Draeger Globe-Trotter and Globe-Trotter IC, Model numbers MS20508 and MS20509. neonatal transport system

Recall number
Z-1365-2022
Recalling firm
Draeger Medical Systems, Inc.
Product
Draeger Globe-Trotter and Globe-Trotter IC, Model numbers MS20508 and MS20509. neonatal transport system
Status
Open, Classified
Initiated
2022-05-20
Posted
not on file
Terminated
not on file
Reason
The firm has initiated this recall to replace component on devices distributed containing a component within the oxygen supply system that is not rated to the working pressure of the system.
Root cause
Under Investigation by firm
Action
On May 20, 2022, the firm began to send Important Field Safety Notice letters to affected customers. Customers were instructed to avoid using the ventilator portion of the system due to the potential for enriched ambient air rising beyond safe levels until a replacement part can be installed and tested. The firm will send customers a replacement kit and testing which will be completed at no charge to the customer. Draeger Service Representatives will call customers and schedule a service visit to update affected systems.
Quantity
38 OUS (no US distribution)
Distribution
No US distribution. Foreign distribution to BY, CL, DE, ES, GB, HU, IN, IT, KW, MX, SA, SV, TW.
Lot, serial or model codes
Model, country, and serial number:  MS20508: El Salvador DSPN-0003  Mexico DSPM-0003, DSPH-0002 Chile DSPL-0003 Saudi Arabia DSPJ-0003, DSPL-0004, DSPL-0005, DSPL-0006, DSPL-0007, DSPM-0001, DSPM-0002, DSPL-0008, DSPK-0002, DSPH-0003, DSPJ-0004, DSRA-0001 Kuwait DSPJ-0002, DSPK-0001 India DSPN-0001, DSPN-0002, DSRA-0002 Taiwan DSPL-0001, SPL-0002, DSPJ-0001  MS20509: Germany DSPH-0003 Qatar DSPL-0007, DSRB-0001 Spain DSPK-0002, DSPJ-0006, DSPJ-0003, DSPH-0001, DSPJ-0004, DSPK-0001, DSPL-0008 Hungary DSPF-0003 Belarus DSPJ-0005 United Kingdom DSPF-0002 Italy DSPJ-0002
510(k) numbers
["K141565"]
PMA numbers
not on file
Product code
FPL
Firm FEI
2510954
Firm city
Telford
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register