Skip to the content

FDA Medical Device Recalls (openFDA)

Medical Information Technology, Inc.: MEDITECH Bi-Directional interactive Communication Analyzer Interface; Integrates with MEDITECH's Laboratory Information System (LIS) with Other Vendor systems to manage the transfer of patient results and demographic information.

Recall number
Z-1365-2013
Recalling firm
Medical Information Technology, Inc.
Product
MEDITECH Bi-Directional interactive Communication Analyzer Interface; Integrates with MEDITECH's Laboratory Information System (LIS) with Other Vendor systems to manage the transfer of patient results and demographic information.
Status
Terminated
Initiated
2013-05-07
Posted
2013-05-22
Terminated
2014-10-01
Reason
Incorrect transmission of laboratory results.
Root cause
Device Design
Action
Beginning on May 7, 2013, MEDITECH distributed a notification to MAGIC and Client Server customers who have the MEDITECH Bi-Directional Interactive Communication Analyzer Interface in use with Vitros 3600, 4600, or 5600 Analyzer in TEST and/or LIVE. This notification was made via remailed task updates that can be printed by the customer. Task updates sent via e-mail are immediately transmitted to the customer. This notification informs customers of the potential patient safety issue that could result in reporting inaccurate qualitative results. Customers will receive a corrective change, in which the previous test's results will no longer be associated with the test not performed. Until customers receive the corrective change, they are instructed to monitor all No Result values from their analyzer and confirm that the result in MEDITECH matches. For further questions please call 781-821-3000, ext. 5742.
Quantity
56 devices
Distribution
Worldwide Distribution - USA, including Puerto Rico and the country of Canada.
Lot, serial or model codes
Magic Releases 5.41, 5.52. 5.53, 5.62, 5.63, 5.64 and Client Server Releases 5.64, 5.65, 6.05, 6.06, 6.07
510(k) numbers
not on file
PMA numbers
not on file
Product code
OUG
Firm FEI
1000120736
Firm city
Westwood
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register