FDA Medical Device Recalls (openFDA)
Fresenius Medical Care Holdings, Inc.: AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US), AquaBplus HF -Typ USA (G02040120FMC-US)
- Recall number
- Z-1364-2022
- Recalling firm
- Fresenius Medical Care Holdings, Inc.
- Product
- AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US), AquaBplus HF -Typ USA (G02040120FMC-US)
- Status
- Terminated
- Initiated
- 2021-08-03
- Posted
- not on file
- Terminated
- 2024-04-10
- Reason
- AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation
- Root cause
- Labeling Change Control
- Action
- Customers were notified via telephone beginning 8/3/21. An FMCRTG, LLC Water Systems Specialist contacted the two consignees via phone and scheduled a time to perform the inspection on the AquaBplus 2000 devices. No formal written communication will be sent since calls and inspections have already performed.
- Quantity
- 2 units
- Distribution
- US Nationwide distribution in the states of NC and WV.
- Lot, serial or model codes
- UDI-DI: (01)04251285200227, (01)04251285200289, (01)0425128520; Serial Numbers: 7BPS1621 7BDS0783 7BFS0517 7BPS1611 7BDS0776 7BFS0502
- 510(k) numbers
- ["K133829"]
- PMA numbers
- not on file
- Product code
- FIP
- Firm FEI
- 3001451489
- Firm city
- Waltham
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28