Skip to the content

FDA Medical Device Recalls (openFDA)

Exactech, Inc.: Truliant Splined Stem Extension, 18mm x 120mm - Product Usage: The devices are used in a cemented total knee prosthesis system, which provides femoral components, tibial inserts, stems, augments, and surgical instrumentation for use in primary and revision total knee arthroplasty.

Recall number
Z-1364-2020
Recalling firm
Exactech, Inc.
Product
Truliant Splined Stem Extension, 18mm x 120mm - Product Usage: The devices are used in a cemented total knee prosthesis system, which provides femoral components, tibial inserts, stems, augments, and surgical instrumentation for use in primary and revision total knee arthroplasty.
Status
Open, Classified
Initiated
2020-01-27
Posted
not on file
Terminated
not on file
Reason
These Truliant Splined Stem Extension, 18mm x 120mm were found to have a labeling error. The upper right corner of the label incorrectly identifies the implant size as 17mm x 80mm when the implant size is 18mm x 120mm.
Root cause
Labeling mix-ups
Action
Exactech distributed the recall notice to the consignees in possession of affected inventory by electronic mail on 7 January 2020. In order to comply with the notice, consignees have been instructed to: "Immediately cease distribution or use of this product. "Extend the information to their accounts that may have this product in their possession. "Identify and quarantine any of the subject devices in their inventory "Complete and return the attached Recall Acknowledgment Form and Recall Inventory Response Form to Exactech.
Quantity
4 units
Distribution
US Nationwide distribution in the states of AZ, CO and FL.
Lot, serial or model codes
Catalog no: 02-012-64-1812  Serial number: 6366405, 6367075, 6367076
510(k) numbers
["K150890"]
PMA numbers
not on file
Product code
JWH
Firm FEI
1038671
Firm city
Gainesville
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register