FDA Medical Device Recalls (openFDA)
Ortho-Clinical Diagnostics: VITROS XT 7600 Integrated System, with V3.4 or 3.4.1 SW Product Code: 6844461 Product Usage: INTENDED USE: For in vitro diagnostic use only. The VITROS XT 7600 Integrated System is intended for use in the measurement of a variety of analytes of clinical interest
- Recall number
- Z-1364-2019
- Recalling firm
- Ortho-Clinical Diagnostics
- Product
- VITROS XT 7600 Integrated System, with V3.4 or 3.4.1 SW Product Code: 6844461 Product Usage: INTENDED USE: For in vitro diagnostic use only. The VITROS XT 7600 Integrated System is intended for use in the measurement of a variety of analytes of clinical interest
- Status
- Terminated
- Initiated
- 2019-03-20
- Posted
- not on file
- Terminated
- 2020-12-08
- Reason
- Potential for sample fluid to be dispensed to an incorrect position on the MicroSlide, potentially leading to erroneous assay results being reported
- Root cause
- Software design
- Action
- Ortho Clinical Diagnostics issued on 20 March 2019, a customer letter via FedEx overnight courier and/ or ORTHO PLUS e-Communications and and/or US Postal Service Priority Mail (for PO Boxes only) customers who were shipped affected VITROS XT 7600 Systems. Software Version 3.4.2 (MOD 4) will be provided for download or DVD format. Contact Ortho Care" Technical Solutions Center at 1-800-421-3311.
- Quantity
- 85 units: US=19;OUS=66
- Distribution
- Worldwide Distribution - US Nationwide in the states of AZ, CA, IL, OH, FL, TX, MI Foreign: Australia, Belgium, Chile, India, Italy, Japan, South Korea, Spain, Canada,, Phillippines
- Lot, serial or model codes
- Serial Number 76000108 thru current.
- 510(k) numbers
- ["K182063"]
- PMA numbers
- not on file
- Product code
- JJE
- Firm FEI
- 1000136573
- Firm city
- Rochester
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28