FDA Medical Device Recalls (openFDA)
Biomet, Inc.: Ziploop Button Product Usage: The Ziploop Button is a non-resorbable system intended to aid in ACL reconstructive procedure due to injury or degenerative disease
- Recall number
- Z-1364-2015
- Recalling firm
- Biomet, Inc.
- Product
- Ziploop Button Product Usage: The Ziploop Button is a non-resorbable system intended to aid in ACL reconstructive procedure due to injury or degenerative disease
- Status
- Terminated
- Initiated
- 2015-02-26
- Posted
- 2015-04-02
- Terminated
- 2016-01-12
- Reason
- The firm is recalling ToggleLocs and Ziploops manufactured in May and July of 2014 as they were not processed following a standard work process, which resulted in complaints of sterile packaging not being sealed.
- Root cause
- Employee error
- Action
- Biomet sent an Urgent Medical Device Recall Notice letter dated March 4, 2015 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediate locate and discontinued use of the product and its return to Biomet. Customers may contact Field Action Specialist, Regulatory Compliance,at (574) 372-1570.
- Quantity
- 50 Units
- Distribution
- Worldwide Distribution: US (nationwide) Distribution in the states of:TX, GA, NM, NC and OK and the countries of: Japan, and the Netherlands.
- Lot, serial or model codes
- PN: 110010873 Lot: 209340
- 510(k) numbers
- ["K080088","K101063"]
- PMA numbers
- not on file
- Product code
- JDR
- Firm FEI
- 1825034
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28