Biomet, Inc.: Ziploop Button Product Usage: The Ziploop Button is a non-resorbable system intended to aid in ACL reconstructive procedure due to injury or degenerative disease

Recall number
Z-1364-2015
Recalling firm
Biomet, Inc.
Product
Ziploop Button Product Usage: The Ziploop Button is a non-resorbable system intended to aid in ACL reconstructive procedure due to injury or degenerative disease
Status
Terminated
Initiated
2015-02-26
Posted
2015-04-02
Terminated
2016-01-12
Reason
The firm is recalling ToggleLocs and Ziploops manufactured in May and July of 2014 as they were not processed following a standard work process, which resulted in complaints of sterile packaging not being sealed.
Root cause
Employee error
Action
Biomet sent an Urgent Medical Device Recall Notice letter dated March 4, 2015 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediate locate and discontinued use of the product and its return to Biomet. Customers may contact Field Action Specialist, Regulatory Compliance,at (574) 372-1570.
Quantity
50 Units
Distribution
Worldwide Distribution: US (nationwide) Distribution in the states of:TX, GA, NM, NC and OK and the countries of: Japan, and the Netherlands.
Lot, serial or model codes
PN: 110010873 Lot: 209340
510(k) numbers
["K080088","K101063"]
PMA numbers
not on file
Product code
JDR
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register