Skip to the content

FDA Medical Device Recalls (openFDA)

Cordis US Corp: PALMAZ GENESIS Transhepatic Biliary Stent on OPTA PRO .035" Delivery System, Catalog numbers PG2990BPS, PG2990BPX, PG3990BPS, PG3990BPX.

Recall number
Z-1363-2022
Recalling firm
Cordis US Corp
Product
PALMAZ GENESIS Transhepatic Biliary Stent on OPTA PRO .035" Delivery System, Catalog numbers PG2990BPS, PG2990BPX, PG3990BPS, PG3990BPX.
Status
Open, Classified
Initiated
2022-05-20
Posted
not on file
Terminated
not on file
Reason
Potential for stent dislodgement and associated failures related to two specific sizes of the device.
Root cause
Process design
Action
The recalling firm issued letters dated 5/16/2022 on 5/20/2022 via FedEx to the U.S. customers. The letter explained the issue, provided details on the affected device to assist in the identification of the product involved, and requested the consignee immediately check their inventory for the affected product and quarantine it. An Acknowledgment Form was enclosed for completion and return to the recalling firm. The consignee was to share the letter with others in their facility who need to be made aware of the recall and with any other facility that may have been sent the affected units.
Quantity
1,089 units
Distribution
US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, WA, WI, and WY. There was also government/military distribution.
Lot, serial or model codes
Catalog #PG2990BPS - Lot numbers 82184806, exp. 12/31/2022; 82191321, exp. 3/31/2023; and 82208532, exp. 10/31/2023. Catalog #PG2990BPX - Lot numbers 82185924, exp. 1/31/2023; and 82208528, exp. 10/31/2023. Catalog #PG3990BPS - Lot numbers 82180267, exp. 10/31/2022; 82191527, exp. 3/31/2023; 82193089, exp. 5/31/2023; and 82206059, exp. 10/31/2023. Catalog #PG3990BPX - Lot numbers 82208524, exp. 11/30/2023; and 82211296, exp. 12/31/2023.
510(k) numbers
["K033394"]
PMA numbers
not on file
Product code
FGE
Firm FEI
1016427
Firm city
Miami Lakes
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register