Philips Medical Systems (Cleveland) Inc: Brilliance 64 and Ingenuity CT Computed Tomography X-Ray Systems. Brilliance 64 Model# 728231, Ingenuity CT Model # 728326, Distributed By: Philips Healthcare, Cleveland, OH. The Brilliance 64 and Ingenuity CT are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories.

Recall number
Z-1361-2012
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Brilliance 64 and Ingenuity CT Computed Tomography X-Ray Systems. Brilliance 64 Model# 728231, Ingenuity CT Model # 728326, Distributed By: Philips Healthcare, Cleveland, OH. The Brilliance 64 and Ingenuity CT are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
Status
Terminated
Initiated
2012-03-12
Posted
2012-04-02
Terminated
2013-06-18
Reason
Philips Healthcare is performing an upgrade to the field for issues related to Philips Brilliance 64 and Ingenuity CT scanner having Software Version 3.5.1. When the user plans a scan with tilt applied, the scan increment changes from the user defined values (noncontiguous) to a contiguous scan.
Root cause
Software design
Action
.
Quantity
440 units
Distribution
Worldwide Distribution-USA (nationwide) including the states of AK, AR, AZ, CA, CO, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MT, ND, NE, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA and WV, and the countries of Australia, Bahamas, Bangladesh, Belgium, Bolivia, Brazil, Chile, China, Colombia, Croatia, Denmark, Finland, France, Georgia, Germany, Great Britain, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Korea, Laos, Lebanon, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Poland, Russia, Slovakia, South Africa, Spain, Sudan, Sweden, Switzerland, Syria, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom and Vietnam.
Lot, serial or model codes
System Code #728231, 4003, 4021, 9033, 9037, 9041, 9045, 9048, 9062, 9064, 9070, 9075, 9080, 9108, 9120, 9142, 9151, 9196, 9197, 9201, 9202, 9205, 9206, 9208, 9214-9218, 9462, 9508, 9531, 9534, 9535, 9544, 9554, 9561, 9584, 9603, 9606, 9608, 9613, 9627, 9630, 9657, 9668, 9671, 9672, 9695, 9706, 9708, 9719, 9728, 9734, 9735, 9741, 9745, 9756, 9789, 9819, 9832, 9838, 9839, 9840, 9859, 9874, 9875, 9891, 9898, 9904, 9918, 9926, 9952, 9964, 9965, 9968, 9975, 9997, 9999, 10000, 10008, 10015, 10041, 10052, 10054, 10092, 10095, 10100, 10103, 10114, 10125, 10145, 10151, 10152, 10154, 10170-10172, 10192, 10196, 10198, 10221, 10222, 10251, 10256, 10273, 10286, 10294, 10307, 10312, 10336, 10337, 10350, 10352, 10362, 10363, 10372, 10376, 10380, 10386, 10393, 10412, 10425, 10426, 10428, 10438, 10445, 10446, 10452, 10459, 10461, 10468, 10470, 10471, 10473, 10482, 10486, 10489, 10491, 10493, 10497, 10503, 10505, 10511-105531, 10533-10555, 10557-10566, 10568-10575, 10577-10604, 10606, 10608-10620, 10622-10624, 10628-10640, 29037, 29049, 29057, 64040, 90036, 90054, 90058, 90063, 90080, 90100, 90102, 90106, 90117, 90135, 90169, 90182, 91003, 95015, 95055, 95065, 95074, 95077, 95101, 95108, 95130, 95133, 95138, 95140, 95157, 95168, 95174, 95178, 95184, 95185, 95191, 95207, 95243, 95295, 95315, 95324, 95327, 95350, 95357, 95358, 95362, 95389, 95403, 95414, 95425, 95429, 95443, 95455, 95504, 95516, 95529, 95534, 95568, 95570, 95572, 95573, 95575, 95602, 95605, 95609, 95613, 95624, 95630, 95638, 95652, 95659, 95664, 95665, 95677, 95684, 95687, 95697, 95700-95728 & 300074.  System Code #: 728326, Serial #: 9511, 300003-300005, 300010-300022, 300024, 300025, 300027-300066 & 300068-300073.
510(k) numbers
["K033326"]
PMA numbers
not on file
Product code
JAK
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register