Philips Medical Systems (Cleveland) Inc: Brilliance 64 and Ingenuity CT Computed Tomography X-Ray Systems. Brilliance 64 Model# 728231, Ingenuity CT Model # 728326, Distributed By: Philips Healthcare, Cleveland, OH. The Brilliance 64 and Ingenuity CT are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
- Recall number
- Z-1361-2012
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Product
- Brilliance 64 and Ingenuity CT Computed Tomography X-Ray Systems. Brilliance 64 Model# 728231, Ingenuity CT Model # 728326, Distributed By: Philips Healthcare, Cleveland, OH. The Brilliance 64 and Ingenuity CT are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
- Status
- Terminated
- Initiated
- 2012-03-12
- Posted
- 2012-04-02
- Terminated
- 2013-06-18
- Reason
- Philips Healthcare is performing an upgrade to the field for issues related to Philips Brilliance 64 and Ingenuity CT scanner having Software Version 3.5.1. When the user plans a scan with tilt applied, the scan increment changes from the user defined values (noncontiguous) to a contiguous scan.
- Root cause
- Software design
- Action
- .
- Quantity
- 440 units
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of AK, AR, AZ, CA, CO, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MT, ND, NE, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA and WV, and the countries of Australia, Bahamas, Bangladesh, Belgium, Bolivia, Brazil, Chile, China, Colombia, Croatia, Denmark, Finland, France, Georgia, Germany, Great Britain, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Korea, Laos, Lebanon, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Poland, Russia, Slovakia, South Africa, Spain, Sudan, Sweden, Switzerland, Syria, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom and Vietnam.
- Lot, serial or model codes
- System Code #728231, 4003, 4021, 9033, 9037, 9041, 9045, 9048, 9062, 9064, 9070, 9075, 9080, 9108, 9120, 9142, 9151, 9196, 9197, 9201, 9202, 9205, 9206, 9208, 9214-9218, 9462, 9508, 9531, 9534, 9535, 9544, 9554, 9561, 9584, 9603, 9606, 9608, 9613, 9627, 9630, 9657, 9668, 9671, 9672, 9695, 9706, 9708, 9719, 9728, 9734, 9735, 9741, 9745, 9756, 9789, 9819, 9832, 9838, 9839, 9840, 9859, 9874, 9875, 9891, 9898, 9904, 9918, 9926, 9952, 9964, 9965, 9968, 9975, 9997, 9999, 10000, 10008, 10015, 10041, 10052, 10054, 10092, 10095, 10100, 10103, 10114, 10125, 10145, 10151, 10152, 10154, 10170-10172, 10192, 10196, 10198, 10221, 10222, 10251, 10256, 10273, 10286, 10294, 10307, 10312, 10336, 10337, 10350, 10352, 10362, 10363, 10372, 10376, 10380, 10386, 10393, 10412, 10425, 10426, 10428, 10438, 10445, 10446, 10452, 10459, 10461, 10468, 10470, 10471, 10473, 10482, 10486, 10489, 10491, 10493, 10497, 10503, 10505, 10511-105531, 10533-10555, 10557-10566, 10568-10575, 10577-10604, 10606, 10608-10620, 10622-10624, 10628-10640, 29037, 29049, 29057, 64040, 90036, 90054, 90058, 90063, 90080, 90100, 90102, 90106, 90117, 90135, 90169, 90182, 91003, 95015, 95055, 95065, 95074, 95077, 95101, 95108, 95130, 95133, 95138, 95140, 95157, 95168, 95174, 95178, 95184, 95185, 95191, 95207, 95243, 95295, 95315, 95324, 95327, 95350, 95357, 95358, 95362, 95389, 95403, 95414, 95425, 95429, 95443, 95455, 95504, 95516, 95529, 95534, 95568, 95570, 95572, 95573, 95575, 95602, 95605, 95609, 95613, 95624, 95630, 95638, 95652, 95659, 95664, 95665, 95677, 95684, 95687, 95697, 95700-95728 & 300074. System Code #: 728326, Serial #: 9511, 300003-300005, 300010-300022, 300024, 300025, 300027-300066 & 300068-300073.
- 510(k) numbers
- ["K033326"]
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 1525965
- Firm city
- Cleveland
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28