Abbott Laboratories: ARCHITECT STAT High Sensitive Troponin-I Controls - Product Usage: The ARCHITECT STAT High Sensitive Troponin-I Controls are for the estimation of test precision and the detection of systematic analytical deviations of the ARCHITECT i System with STAT protocol capability when used for the quantitative determination of cardiac troponin I (cTnI) in human serum and plasma with the ARCHITECT STAT High Sensitive Troponin-I reagent kit.

Recall number
Z-1359-2019
Recalling firm
Abbott Laboratories
Product
ARCHITECT STAT High Sensitive Troponin-I Controls - Product Usage: The ARCHITECT STAT High Sensitive Troponin-I Controls are for the estimation of test precision and the detection of systematic analytical deviations of the ARCHITECT i System with STAT protocol capability when used for the quantitative determination of cardiac troponin I (cTnI) in human serum and plasma with the ARCHITECT STAT High Sensitive Troponin-I reagent kit.
Status
Terminated
Initiated
2019-04-09
Posted
not on file
Terminated
2019-12-20
Reason
Devices were delivered without the required dry ice.
Root cause
Under Investigation by firm
Action
On about 04/04/2019, Abbott notified customers in Latvia of the recall. Instructions included to examine inventory for any affected devices on hand, review the letter with your Medical Director or Laboratory Management and follow your laboratory protocol regarding the need for reviewing previously reported patient results, complete and return the Customer Reply form, and notify customers if the affected devices have been forwarded to other laboratories and provide them a copy of the notification letter. If replacement material is available, customers were instructed to discontinue the use of affected devices and destroy any remaining inventory according to your laboratory procedures. If replacement material is not available, immediately order replacement material. Abbott instructed customers that they may continue to use the impacted material as long as controls are within range. Destroy impacted material once replacement material is available.
Quantity
2 units
Distribution
Distribution to Latvia.
Lot, serial or model codes
LN 3P25-11; Lot 94347UIOO
510(k) numbers
not on file
PMA numbers
not on file
Product code
JIT
Firm FEI
1415939
Firm city
Abbott Park
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register