FDA Medical Device Recalls (openFDA)
Zimmer Surgical Inc: Zimmer Sterile Disposable Tourniquet Cuff with Protective Sleeve and PLC, Dual Port, Dual Bladder, Item No. 60708015200, UDI (01)00889024377547 (17)230812(10)27968701 The product is a tourniquet - Product Usage: Tourniquets are intended to be used by qualified medical professionals to exert enough pressure on the arterial blood flow in a limb to produce a bloodless operating field.
- Recall number
- Z-1356-2020
- Recalling firm
- Zimmer Surgical Inc
- Product
- Zimmer Sterile Disposable Tourniquet Cuff with Protective Sleeve and PLC, Dual Port, Dual Bladder, Item No. 60708015200, UDI (01)00889024377547 (17)230812(10)27968701 The product is a tourniquet - Product Usage: Tourniquets are intended to be used by qualified medical professionals to exert enough pressure on the arterial blood flow in a limb to produce a bloodless operating field.
- Status
- Terminated
- Initiated
- 2020-01-03
- Posted
- 2020-01-28
- Terminated
- 2020-10-30
- Reason
- When inflating/deflating a single bladder, both bladders could potentially inflate/deflate due to a leak between the bladders.
- Root cause
- Process control
- Action
- On January 3, 2020, the recalling firm distributed recall notices to affected customers via email and FedEx. Customers were instructed to ensure that affected personnel are aware of the contents, and assist the Zimmer Biomet sales representative to locate, quarantine, and return all affected product. If you have questions or concerns regarding this recall, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com.
- Quantity
- 58
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, FL, IA, ID, IL, LA, MA, MO, NC, NY, OH, OR, PA, SC, TN, TX, WA, WI, and WV and country of Netherlands.
- Lot, serial or model codes
- Lot 27968701
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- KCY
- Firm FEI
- 1526350
- Firm city
- Dover
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as