Skip to the content

FDA Medical Device Recalls (openFDA)

Cook Inc.: Zenith Alpha Spiral-Z Endovascular Leg, Catalog Prefix ZISL. The product is an endovascular graft.

Recall number
Z-1355-2020
Recalling firm
Cook Inc.
Product
Zenith Alpha Spiral-Z Endovascular Leg, Catalog Prefix ZISL. The product is an endovascular graft.
Status
Terminated
Initiated
2020-01-16
Posted
2020-01-31
Terminated
2020-11-12
Reason
The firm has identified factors which contributed to failures relating to thrombus formation and/or lumen occlusion reports, and is highlighting to customers key points of the IFU pertaining to the prevention of these factors. In addition, the firm has updated the Planning and Sizing Worksheet to include associated information.
Root cause
Labeling design
Action
On January 14, 2020, the firm notified affected customers via Urgent Medical Device Correction letters/emails. Customers were advised to pay especial attention to several aspects of the Instructions for Use (IFU) which were of key importance when implanting a device. Customers were also provided with an updated Planning and Sizing Worksheet. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.
Quantity
68,241
Distribution
Foreign distribution only. Distributed worldwide.
Lot, serial or model codes
All lots. OUS distribution only; Product is not cleared or distributed in the USA.
510(k) numbers
not on file
PMA numbers
not on file
Product code
MIH
Firm FEI
1820334
Firm city
Bloomington
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register