Integra LifeSciences Corp.: CODMAN HAKIM BACTISEAL Shunt System Clear with Barium Stripe Peritoneal Catheter 120cm, with drainage Holes and Marking. For use in the treatment of hydrocephalus. Catalog Number: NS5524

Recall number
Z-1354-2021
Recalling firm
Integra LifeSciences Corp.
Product
CODMAN HAKIM BACTISEAL Shunt System Clear with Barium Stripe Peritoneal Catheter 120cm, with drainage Holes and Marking. For use in the treatment of hydrocephalus. Catalog Number: NS5524
Status
Terminated
Initiated
2021-03-17
Posted
2021-04-05
Terminated
2021-12-02
Reason
Incorrect labeling, Catheter did not have the required holes and markings
Root cause
Labeling Change Control
Action
Integra LifeSciences issued Voluntary Urgent Medical Device Recall letter via courier service, facsimile or email on March 17th for delivery on March 19th, 2021. Letter states reason for recall, health risk and action to take: If you do have the affected product in Table 1, remove the product from service. b. Complete the attached acknowledgement form. If you do have affected product, check the box I do have affected product. Record the total quantity of the affected product you have. c. If you do not have affected product, check the box, I do not have affected product. d. Return the completed acknowledgement form to FCA2@integralife.com or FAX to 1-609-750-4220 e. Keep a copy of the form for your records. f. When your form is received, and it is noted that you have the affected product, Customer Service will contact you and provide a Return Material Authorization (RMA) number and directions to return the product. A replacement order will also be placed for the quantity noted on the form. Receipt of the acknowledgement form ensures that Integra has achieved a level of effectiveness in communicating this information effectively. We recommend you also maintain a copy of this notification and signed copy of the acknowledgement forms for your records. Regulatory agencies such as the FDA perform audits of field actions of this nature to verify that our customers have been notified and understand the nature of the field action. Should you have any questions regarding these instructions, please contact Customer Service: USA: 800-654-2873 Monday- Friday 8:00AM EST 6:00PM EST custsvcnj@integralife.com
Quantity
49 units
Distribution
CA CO FL KY MD NC NC NY NY TX WI
Lot, serial or model codes
Lot Number: 5028205  UDI: (01)10381780526797 (17)210930(10)5028205
510(k) numbers
["K031123"]
PMA numbers
not on file
Product code
HCA
Firm FEI
3003418325
Firm city
Princeton
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register