FDA Medical Device Recalls (openFDA)
Abbott Ireland Diagnostics Division: Abbott ARCHITECT Estradiol Reagent Kit - Product Usage: The ARCHITECT Estradiol assay is a Chemiluminescent Microparticle Immunoassay (CMIA) for the quantitative determination of Estradiol in human serum and plasma.
- Recall number
- Z-1353-2019
- Recalling firm
- Abbott Ireland Diagnostics Division
- Product
- Abbott ARCHITECT Estradiol Reagent Kit - Product Usage: The ARCHITECT Estradiol assay is a Chemiluminescent Microparticle Immunoassay (CMIA) for the quantitative determination of Estradiol in human serum and plasma.
- Status
- Terminated
- Initiated
- 2019-02-05
- Posted
- not on file
- Terminated
- 2020-08-21
- Reason
- Patient results may be falsely elevated. This patient impact only applies to patients currently being treated with or recently treated with the drug Mifepristone.
- Root cause
- Under Investigation by firm
- Action
- On February 5, 2019, the firm sent Product Correction Urgent - Immediate Action Required letters to all their consignees which alerted them to the current issue. We recommend that you follow the necessary actions below: Patients treated with Mifepristone should not be tested with the ARCHITECT or Alinity i Estradiol assay for up to two weeks based on information available regarding Mifepristones bioavailability. - Please review this letter with your Medical Director. - If you have forwarded the products listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of this letter. - Complete the customer reply form. - Please retain this letter for your laboratory records. If you or any of the health care providers you serve have any questions regarding this information, U.S. Customers please contact Customer Service at 18774ABBOTT (available 24 hours a day, 7 days a week). Customers outside the U.S., please contact your local area Customer Service.
- Quantity
- 2,724
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR. AZ, CA, CT, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, KS, MT, NC, ND, NJ, NV, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA, WI, AND PUERTO RICO, and countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Azerbaijan, Belarus, Belgium, Bosnia and Herzegovi, Brazil, Brunei, Botswana, Canada, Chile, China, Columbia, Costa Rica, Croatia, Curacao, Cyprus, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Gaza and Jericho, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland,Israel, Italy, Jamaica, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Maldives, Mali, Mauritius, Mexico, Moldova, Montenegro, Morocco, Mozambique, Namibia, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland,Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turks, UAE, Uganda, United Kingdom, and Vietnam.
- Lot, serial or model codes
- List Number 7K72-20 - ALL LOTS, List Number 7K72-25 - ALL LOTS, and List number 7K72-35 - ALL LOTS.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- CHP
- Firm FEI
- 3005094123
- Firm city
- Lisnamuck Co. Longford
- Firm state
- not on file
- Firm country
- Ireland
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28