FDA Medical Device Recalls (openFDA)

Philips Medical Systems (Cleveland) Inc: Brilliance 64:Computed Tomography X-ray Systems, Intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes

Recall number
Z-1351-2016
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Brilliance 64:Computed Tomography X-ray Systems, Intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes
Status
Terminated
Initiated
2016-02-05
Posted
not on file
Terminated
2018-07-02
Reason
The firm was notified of a software error in which the system may not automatically send all image/data series to remote devices.
Root cause
Software design
Action
Philips Medical Systems sent a Customer Information dated February 5, 2016, to all affected customers via certified mail. The letter identified the product, the problem, and the action to be taken by the customer. To correct the issue Philips will release field change orders (FCO) 72800652 and 72800644 to install a software update on the affected systems. Customers with questions were instructed to contact their local Philips representative or local Philips Healthcare office. For North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377, Option 5: Enter Site ID or follow the prompts). For questions regarding this recall call 440-483-2997.
Quantity
285 units
Distribution
Worldwide Distribution - US to the District of Columbia: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA & WV. Product was also shipped to the following countries: Afghanistan, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, China, Cuba, Denmark, Egypt, Ethiopia, Finland, France, Germany, Ghana, Gibraltar, Greece, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kyrgyzstan, Lebanon, Lithuania, Macedonia, Malawi, Mexico, Myanmar, Netherlands, New Zealand, Norway, Oman, Palestinian, Philippines, Poland, Republic of Korea, R¿union, Romania, Russian Federation, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan & Viet Nam.
Lot, serial or model codes
Model No. 728231; S/N: 1, 2, 4001, 9056, 9074, 9099, 9134, 9251, 9515, 9529, 9538, 9549, 9554, 9573, 9574, 9618, 9625, 9638, 9646, 9653, 9656, 9659, 9667, 9670, 9686, 9689, 9690, 9703, 9738, 9740, 9755, 9782, 9811, 9822, 9833, 9837, 9838, 9859, 9862, 9898, 9901, 9909, 9910, 9917, 9920, 9923, 9927, 9930, 9933, 9983, 9989, 9993, 10021, 10024, 10042, 10047, 10050, 10051, 10077, 10092, 10099, 10106, 10150, 10180, 10189, 10193, 10206, 10206, 10208, 10213, 10250, 10253, 10299, 10379, 10394, 10422, 10670, 10671, 10677, 10678, 10683, 10684, 10685, 10695, 10698, 10794, 10799, 10803, 29002, 29014, 29056, 29070, 29107, 29115, 90100, 90124, 90150, 90154, 91003, 95022, 95024, 95100, 95130, 95157, 95191, 95197, 95213, 95222, 95236, 95258, 95261, 95333, 95351, 95364, 95387, 95399, 95410, 95473, 95475, 95496, 95504, 95507, 95512, 95524, 95551, 95563, 95614, 95616, 95668, 95688, 95691, 95703, 95704, 95705, 95706, 95707, 95708, 95709, 95710, 95711, 95713, 95716, 95717, 95718, 95720, 95721, 95722, 95723, 95724, 95725, 95726, 95727, 95728, 95729, 95730, 95731, 95732, 95733, 95734, 95735, 95736, 95737, 95738, 95739, 95740, 95741, 95742, 95744, 95745, 95746, 95747, 95748, 95749, 95750, 95751, 95752, 95753, 95754, 95755, 95756, 95757, 95758, 95759, 95760, 95761, 95762, 95764, 95765, 95766, 95767, 95769, 95770, 95772, 95773, 95774, 95775, 95776, 95777, 95778, 95779, 95780, 95781, 95786, 95787, 95788, 95789, 95790, 95791, 95794, 95795, 95797, 95798, 95799, 95800, 95801, 95802, 95803, 95804, 95805, 95806, 95807, 95808, 95809, 95810, 95811, 95812, 95813, 95815, 95817, 95821, 95824, 95830, 95831, 95832, 95834, 95835, 95837, 95842, 95843, 95845, 95849, 95852, 95853, 95854, 95855, 95857, 95864, 95869, 95874, 95876, 95877, 95881, 95886, 95900, 95901, 95902, 95903, 95904, 95906, 95907, 95910, 95911, 95912, 95915, 95916, 95917, 95918, 95919, 95922, 95924, 95925, 95927, 95928, 95929, 95930, 95934, 95938, 95940, 95941, 95942, 95944, 95946, 95341A, 123002 & 123029.
510(k) numbers
["K033326"]
PMA numbers
not on file
Product code
JAK
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register