FDA Medical Device Recalls (openFDA)
Medical Components, Inc dba MedComp: Medcomp Duo-Flow 400XL Catheter, Hemo-Flow Catheter, SLX Catheter, Triple Lumen Infusion Catheter, CT rated PICC, Split Cath III, PC Split Cath III; AngioDynamics product- Dura-Flow Catheter, Schon XL Catheter; The affected product are insertion kits used to implant the devices packaged-long and short term hemodialysis catheters and infusion catheters.
- Recall number
- Z-1351-2015
- Recalling firm
- Medical Components, Inc dba MedComp
- Product
- Medcomp Duo-Flow 400XL Catheter, Hemo-Flow Catheter, SLX Catheter, Triple Lumen Infusion Catheter, CT rated PICC, Split Cath III, PC Split Cath III; AngioDynamics product- Dura-Flow Catheter, Schon XL Catheter; The affected product are insertion kits used to implant the devices packaged-long and short term hemodialysis catheters and infusion catheters.
- Status
- Terminated
- Initiated
- 2015-02-05
- Posted
- 2015-03-30
- Terminated
- 2015-08-06
- Reason
- Medcomp has initiated the recall of Duo-Flow 400XL Catheter, Hemo-Flow Catheter, SLX Catheter, Triple Lumen Infusion Catheter, CT rated PICC, Split Cath III, PC Split Cath III because the product shipped was out of specification. One BI (biological indicator) out of 21 on the sterilization load was out of specification.
- Root cause
- Process control
- Action
- On or about 02/03/2015, Medcomp sales representatives notified the affected medical facilities and providing Medcomp's recall letter. This was accomplished by e-mail, telephone, or visit. Distributors for the international accounts were notified by Medcomp via e-mail. A copy of Medcomp's recall letter was attached. For AngioDynamics product- Medcomp notified AngioDynamics by e-mail of the product recall. Every consignee was requested to complete and return Page 2 of the recall letter verifying receipt of the recall and disposition of the product they received. AngioDynamics was requested to provide an account of the disposition for the total number of product they received.
- Quantity
- 3152
- Distribution
- Worldwide Distribution.
- Lot, serial or model codes
- Medcomp products:Catalog Code/ Lot Number/Expiration Date: DFXL148MTE/ MGMC610/10/31/2019; HFS 32/ MGMB440/ 07/14/2017; HFS24E./MGLY830/07/31/2017; MC061402/MGMD190/07/15/2017; MC3L-8S/MGMD920/10/16/2019; MR17035211/MGMD680/10/31/2017; TRAY #593-3/MGMB460/07/15/2017; RMS23602/MBWM930/07/31/2017; ASPC2816-3/MBWM900/07/31/2017; Tray #55216-3/MBWL780/07/31/2017; ASPC2816-3PC/MBWX290/07/31/2017; 10301207/MGMD930/07/31/2017; 10800701/ MGLY730/ 07/31/2017; 10800702/ MGMD210/ 07/31/2017; 10800703/MGLY740/07/31/2017 AngioDynamics products- model numbers/lot numbers/ expiration dates: 10301207/MGMD930/07/31/2017; 10800701/MGLY730/07/31/2017; 10800702/MGMD210/07/31/2017; 10800703/MGLY740/07/31/2017
- 510(k) numbers
- ["K030502","K111651","K994105"]
- PMA numbers
- not on file
- Product code
- MSD
- Firm FEI
- 2518902
- Firm city
- Harleysville
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28