ulrich medical USA Inc: Ulrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw-to-screw

Recall number
Z-1350-2020
Recalling firm
ulrich medical USA Inc
Product
Ulrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw-to-screw
Status
Terminated
Initiated
2019-12-31
Posted
not on file
Terminated
2025-11-17
Reason
Due to a production error, the weld seam of the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation.
Root cause
Process control
Action
Ulrich notified initially notified customers, via phone call, on 12/20/2019. On about 12/31/2019, customers were also sent an official "URGENT SAFETY INFORMATION: MEDICAL DEVICE CORRECTIONS & REMOVALS" letter. Customers were informed that the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation. Instructions included cease use of the affected devices, return them to Ulrich, and complete and return the provided "Product Recall" form. Customers were also instructed to inform all users of the devices within each organization are informed. Questions or concerns can be directed to Mr. Louis Milos, Sr. Manager Business and Quality Systems, at 636-519-0268 ext. 105, l.milos@ulrichmedicalusa.com, Monday through Friday 8:00am to 5:00pm CST.
Quantity
179 units
Distribution
AZ, TX, GA, FL, AR, CA
Lot, serial or model codes
Lot Numbers: U010577, U016620, U019084
510(k) numbers
["K150650"]
PMA numbers
not on file
Product code
KWP
Firm FEI
3005532635
Firm city
Chesterfield
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register