Skip to the content

FDA Medical Device Recalls (openFDA)

Randox Laboratories, Limited: RX Daytona Plus (with ISE/without ISE); Models Nos. RX4040 (w/ISE) and RX4041 (without ISE) - Product Usage: A fully automated, random access, clinical chemistry analyzer complete with dedicated analyzer software. For use in clinical laboratories.

Recall number
Z-1349-2020
Recalling firm
Randox Laboratories, Limited
Product
RX Daytona Plus (with ISE/without ISE); Models Nos. RX4040 (w/ISE) and RX4041 (without ISE) - Product Usage: A fully automated, random access, clinical chemistry analyzer complete with dedicated analyzer software. For use in clinical laboratories.
Status
Terminated
Initiated
2019-12-19
Posted
2020-02-24
Terminated
2021-04-23
Reason
Software version UI2550642107 for the RX Daytona + instrument released to correct the following issues: 1) The order of samples displayed is not sequential according to SID (sample identification number) when searching the run/results screen; 2) Incorrect color flagging of IQC as "out of range."
Root cause
Under Investigation by firm
Action
Randox Medical Device Correction letter dated 12.17.2019. The recall from the manufacturer is extended to the distribution center within the USA and Puerto Rico. The distributor will then contact the customer directly. The initial contact and the follow up contact will be via email. If further contacts are required these will be via telephone. It is intended that the initial contact will be conducted 19th December 2019. The distributors and customers will be instructed to read and review the release notes and carry-out all actions specified within. Distributors are advised to contact all end-users and provide the FSN, response form and release notes. Customers were advised to take the following actions: complete and return the response form 12187-QA to technical.services@randox .com within five working days. ***Updated 3/2/2020*** The firm sent an updated customer letter on February 28, 2020, which included an updated "Risk to Health" statement, which now states that a delay in reporting results could lead to a delay in diagnosis and/or treatment. No erroneous patient results are generated due to this issue. Instructions to customers have not changed.
Quantity
U.S.: RX4040-2 units RX 4041-3 units
Distribution
US Nationwide distribution including state of PR.
Lot, serial or model codes
Software version UI2550642107 for the RX Daytona Plus instrument.
510(k) numbers
["K131554"]
PMA numbers
not on file
Product code
JJE
Firm FEI
1000361607
Firm city
Crumlin
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register