GETINGE US SALES LLC: VOLISTA StandOP Surgical Light; Model numbers VLT600 SF AIM STP, VLT600 DF AIM STP, Ref. codes ARD568811901, ARD568811911 ARD568811951, ARD568811961 The device is designed to illuminate the body of a patient during surgical operations, diagnostics and treatment. These surgical lights may be installed in surgical suites, examining rooms, doctors' surgeries and outpatient consultations.

Recall number
Z-1349-2018
Recalling firm
GETINGE US SALES LLC
Product
VOLISTA StandOP Surgical Light; Model numbers VLT600 SF AIM STP, VLT600 DF AIM STP, Ref. codes ARD568811901, ARD568811911 ARD568811951, ARD568811961 The device is designed to illuminate the body of a patient during surgical operations, diagnostics and treatment. These surgical lights may be installed in surgical suites, examining rooms, doctors' surgeries and outpatient consultations.
Status
Terminated
Initiated
2018-02-07
Posted
2018-04-10
Terminated
2021-09-02
Reason
The manufacturer received complaints indicating that the device's central handle holder detaches easily from the handle holder ring on the light head during clinical use. In the event of such detachment, particles from the handle holder ring could fall into the sterile field during a procedure.
Root cause
Component design/selection
Action
The firm, Getinge, sent an "URGENT MEDICAL DEVICE RECALL" Notice dated 2/7/2018 to all affected consignees via mail. The notice described the product, problem and actions to be taken. The consignees were instructed to do the following: 1) Ensure that all users of the device are made aware of this field correction. 2) Check the device serial number against the list of affected serial numbers. 3) Contact the recalling firm to arrange for servicing of all affected units. Getinge Customer Service: (800) 950-9912, 8AM - 5PM (EST) Monday to Friday. 4) Complete and sign the enclosed customer response form, and return it by fax to: (973) 860-1675 or by email to: VolistaLights.FA@getinge.com. If you have any questions, please contact your Getinge representative or call Getinge Customer Service at (800)950-9912 (Select option 1) between 8:00 AM and 5:00 PM (EST).
Quantity
5214
Distribution
Worldwide Distribution: US (nationwide) to states of: AR, AZ, CA, CO, FL, ID, IA, KS, KY, ME, MO, NE, NY, OR, TN, TX, UT, and WA; and countries of: Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Cameroon, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Ghana, Hong Kong, Hungary, India, Iran, Iraq, Italy, Japan, Jordan, Kenya, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Namibia, Netherlands, Nigeria, Oman, Pakistan, Panama, Paraguay, Poland, Portugal, South Korea, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Trinidad-Tobago, Turkey, UAE, UK, Yemen and Zimbabwe.
Lot, serial or model codes
Model VLT600 SF AIM STP:  Ref Code: ARD568811901: Serial Numbers:  21001 21002 22000  Ref Code: ARD568811911: Serial Numbers: 25002 31000 31001 31002 31003 31004 31006 31008 31009 31010 31012 31017 31018 31019 31020 31021 31022 31023 31024 31025 31026 35000 35001 35011 35012 35013 35014 35015 35017 35018 35019 35020 35021 35022 35023 35024 35025 35026 35027 35028 35029 35030 35031 35032 35033 38000 38001 38002 38003 38005 38009 38011 38012 38013 38014 38015 38016 38017 38018 38019 38020 38021 38022 38023 38024 38025 38026 38027 500000 500001 500002 500003 500004 500005  Model VLT600 DF AIM STP:  Ref Code: ARD568811951 (Not in distribution in the U.S.) Serial Numbers: 20502 20503  Ref Code: ARD568811961 Serial Numbers: 25005 25006 25007 28003 28004 31000 31001 31002 31003 31004 31005 31006 31008 31009 31011 31012 31013 31014 31015 31016 31017 31019 31020 31030 31031 31039 31040 31041 31042 31043 31044 31045 31046 31047 31048 31049 31050 31051 31052 31053 31054 31055 31056 38025 38026 38027 38028 500001 500002
510(k) numbers
["K130513"]
PMA numbers
not on file
Product code
FSY
Firm FEI
3012092534
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register