Skip to the content

FDA Medical Device Recalls (openFDA)

Intuitive Surgical, Inc.: Da Vinci S 8mm Instrument Cannula with outlet for use with the Da Vinci Surgical System; Model IS2000, Part Numbers: 400254-01, 400255-01; Used in endoscopic surgery to establish a port of entry for Intuitive Surgical's EndoWrist Instruments. Product is distributed by Intuitive Surgical, Inc., Sunnyvale, CA

Recall number
Z-1348-2008
Recalling firm
Intuitive Surgical, Inc.
Product
Da Vinci S 8mm Instrument Cannula with outlet for use with the Da Vinci Surgical System; Model IS2000, Part Numbers: 400254-01, 400255-01; Used in endoscopic surgery to establish a port of entry for Intuitive Surgical's EndoWrist Instruments. Product is distributed by Intuitive Surgical, Inc., Sunnyvale, CA
Status
Terminated
Initiated
2008-01-04
Posted
2008-08-06
Terminated
2011-03-31
Reason
(1) Incorrect dimension on Luer on smoke evacuation cannulae not allowing for secure attachment function. (2) Incorrect labeling. External labeling lot number differs from lot number etched on the cannula.
Root cause
Process control
Action
On 1/4/08, the firm initiated the recall and its notification was via Urgent Device Recall letters explaining the reason for the recall, requesting immediate discontinued use, segregate non-conforming product in a secure area, and a Customer Service Rep will make contact in order to arrange for the retrieval of the material.
Quantity
39 units
Distribution
Nationwide
Lot, serial or model codes
Lot Numbers: 13730-01-70807, 13793-71106, 13730-02-70807, WI073202, WI074501, WI073201
510(k) numbers
not on file
PMA numbers
not on file
Product code
NAY
Firm FEI
3001675293
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register