Philips North America, LLC: Philips DigiTrak XT Holter Recorder Software version number B.1 and C.2. The DigiTrak XT Recorder is available in four different configurations. Service tag numbers for these configurations are as follows: - 453564067151 24-Hour Configuration - 453564067161 48-Hour Configuration - 453564067171 96-Hour Configuration - 453564067181 7-Day Configuration Model Number: 860322
- Recall number
- Z-1345-2020
- Recalling firm
- Philips North America, LLC
- Product
- Philips DigiTrak XT Holter Recorder Software version number B.1 and C.2. The DigiTrak XT Recorder is available in four different configurations. Service tag numbers for these configurations are as follows: - 453564067151 24-Hour Configuration - 453564067161 48-Hour Configuration - 453564067171 96-Hour Configuration - 453564067181 7-Day Configuration Model Number: 860322
- Status
- Terminated
- Initiated
- 2020-01-13
- Posted
- not on file
- Terminated
- 2023-06-22
- Reason
- if an AAA battery is inserted in the recorder and a user attempts to start it, or if the recorder is inserted in the docking station, the recorder will display Error: 602 and fail to function for a new patient study.
- Root cause
- Device Design
- Action
- Philips Issued Medical Device Correction letter dated January 13, 2020 to U.S. customers, via priority mail. Philips Market Organizations are responsible for distributing the letters outside of the U.S. Instructions to Customer Philips is asking customers to follow the Action to be Taken by Customer/User section of the CIL: Upon receipt of this notification, promptly locate and isolate affected devices. The steps in the Instructions on Clearing Error 602 below may be used to reset your devices. Customers can contact Philips should they choose not to reset the device on their own. For information or support contact: your local Philips representative or call us at 1 -800-722-9377.
- Quantity
- 44,442 units
- Distribution
- US Nationwide Foreign: Afghanistan Albania Algeria Argentina Armenia Aruba Australia Austria Azerbaijan Bangladesh Belarus Belgium Bermuda Bulgaria Burkina Faso Canada China C¿te D'Ivoire Croatia Curacao Cyprus Czech Republic Egypt France Germany Ghana Gibraltar Greece Guatemala Hong Kong Hungary Iceland India Indonesia Iran Iraq Ireland Israel Italy Japan Jordan Kenya Korea, Republic of Kosovo Kuwait Latvia Lebanon Libya Luxembourg Macao Macedonia Malaysia Maldives Mali Malta Mauritius Mexico Mongolia Myanmar Nepal Netherlands Netherlands Antilles New Zealand Nicaragua Nigeria Norway Oman Pakistan Palestine, State of Panama Peru Philippines Poland Portugal Puerto Rico Qatar Romania Russian Federation Rwanda Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Suriname Sweden Switzerland Taiwan Thailand Trinidad and Tobago Turkey Uganda United Arab Emirates United Kingdom Uzbekistan Venezuela Viet Nam
- Lot, serial or model codes
- All Philips 860322 DigiTrak XT Holter Recorder, manufactured or distributed before December 31, 2019, are affected, regardless of serial number, ship date, and date of manufacture.
- 510(k) numbers
- ["K071733"]
- PMA numbers
- not on file
- Product code
- MWJ
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28