FDA Medical Device Recalls (openFDA)
Technidata S.A.: TD-Synergy. Laboratory Information System.
- Recall number
- Z-1345-2013
- Recalling firm
- Technidata S.A.
- Product
- TD-Synergy. Laboratory Information System.
- Status
- Terminated
- Initiated
- 2010-07-29
- Posted
- 2013-05-21
- Terminated
- 2013-12-05
- Reason
- Technidata has discovered that using the F8 key in the ERM session to add a combined test can result in the modification of the value of a result that has already been entered for a test of the request.
- Root cause
- Software design
- Action
- The firm, Technidata, sent an "P1 DEFECT INFORMATION" notice dated July 2, 2010 to their consignee/customers. The notice describes the product, problem and actions to be taken. The notice directs them to find the correction to the software issue on the Technidata website where they are able to install the correction kit depending on the version and operating system used. Note "You must back up the database and the installation directory before installing this correction". Should you need any further information, do not hesitate to contact the Support team at +33 (0) 4 76 04 13 00.
- Quantity
- 10
- Distribution
- Worldwide Distribution-USA (nationwide) including Washington DC and the states of California, Nevada, Kentucky, Alabama, Rhode Island, Pennsylvania, and Louisiana and the country of Italy.
- Lot, serial or model codes
- TD Lims product versions: LIS (ALLX)-21.041-_ML-5370, LIS (ALLX)-22.021-_ML-5374, TD-Synergy product versions: SYN (ALLX)- 03.13B-_ML-5368, SYN (ALLX)- 03.141-_ML-5369, SYN(LINUX)-06.011-_ML-5381, SYN (ALLX)-11.11B-_ML-5345, SYN (ALLX)-11.21B-_ML-5346 SYN (ALL)- 11.31C-_ML-5372, SYN (LINUX)-11.321-_FR-5382, SYN (ALL)-11.321-_ML-5373
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JQP
- Firm FEI
- 3003995437
- Firm city
- Montbonnot St Martin
- Firm state
- not on file
- Firm country
- France
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28