Skip to the content

FDA Medical Device Recalls (openFDA)

Fenwal Inc: Amicus Exchange Kit - Product Usage: intended for use in the collection of blood components and mononuclear cells.

Recall number
Z-1344-2021
Recalling firm
Fenwal Inc
Product
Amicus Exchange Kit - Product Usage: intended for use in the collection of blood components and mononuclear cells.
Status
Terminated
Initiated
2021-02-25
Posted
not on file
Terminated
2023-04-18
Reason
Some lots of Amicus MNC Apheresis and Amicus Exchange kits have leaking centrifuge packs during a procedure on the Amicus Separator using therapeutic protocols only. The defect is a blood leak at the boot, elbow, or in the channel on the separation chamber of the centrifuge pack.
Root cause
Under Investigation by firm
Action
A Recall notification letter titled, "URGENT - Medical Device Recall Amicus MNC Apheresis/Exchange Disposable Kits" and form titled, Amicus Therapeutic Kits Customer Reply Form were sent to consignees on 02/25/2021 via Fed Ex Overnight Express. The consignee requested their customers to: 1) Discontinue use of the affected products immediately by checking their inventory and quarantining all affected product at their facility. They were also informed to contact Fresenius Kabi Customer Service at 800-333-6925, Prompt 1 for replacement kits to ensure additional procedures can continue. 2) Inform potential users of the product in their organization of this notification. 3) Based on the options listed below, complete the attached response form and return via the fax number, email address, or the enclosed Fed Ex envelope and return address. Option A: If there is no remaining inventory of the affected product complete the response form and return it to Fresenius Kabi. Option B: If the customer has affected product in inventory and wishes to return it: Contact Fresenius Kabi Customer Service at 800-333-6925, Prompt 1 and request to return the affected product. They will assist the customer with placing a replacement order at that time. Option C: If the customer has a small number of kits in inventory and are willing to destroy the product on Fresenius Kabis behalf: Contact Fresenius Kabi Customer Service at 800-333-6925, Prompt 1 to identify if the customer has product to be destroyed and require replacement. They will provide a certificate of destruction for completion. Once completed, the certificate should be returned to Fresenius Kabi and replacements will be sent. Customers may be in a position that no alternate Device is available and therefore choose to continue with the use of the affected Amicus MNC Apheresis kits or Amicus Exchange kits (Therapeutics kits). For such cases, the following actions should be put in place until replacement products a
Quantity
2466 units
Distribution
US Nationwide distribution in the states of GA, TX, TN, MD, CA, KS, MN, MA, NM, NC, LA, CO, KY, NV, PA, UT, MI, NJ, IL, MD, OH, NY, FL.
Lot, serial or model codes
Product Code X6R2339; UDI: 04086000100151; Batch Numbers FA20F08147 (Exp. 06/07/2022), FA20H12145 (Exp. 08/11/2022), FA20J27157 (Exp. 10/26/2022)
510(k) numbers
["K180615"]
PMA numbers
not on file
Product code
LKN
Firm FEI
3004548776
Firm city
Lake Zurich
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28