Skip to the content

FDA Medical Device Recalls (openFDA)

ICU Medical, Inc.: 7" Ext Set w/MicroClave, Clave T-Connector, Clamp, Rotating Luer, 50 pouches per case, Item AH7262. The firm name on the label is icumedical, San Clemente, CA. The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).

Recall number
Z-1344-2018
Recalling firm
ICU Medical, Inc.
Product
7" Ext Set w/MicroClave, Clave T-Connector, Clamp, Rotating Luer, 50 pouches per case, Item AH7262. The firm name on the label is icumedical, San Clemente, CA. The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Status
Terminated
Initiated
2018-02-05
Posted
not on file
Terminated
2019-09-17
Reason
The infusion sets were manufactured with an incorrect spin collar.
Root cause
Process control
Action
The firm, ICU Medical Inc., sent an "URGENT: Medical Device Recall Notification" letter dated 2/5/2018 to its customers via UPS on 2/5/2018 in the U.S. The letter described the product, problem and actions to be taken. The customers were instructed to: inspect your inventory for specific product and lot numbers; quarantine all affected devices; complete and return the Recall Response Form as applicable (even if you have no affected devices) via Fax to: 1-866-899-7474 or email to: ICUMedical2805@stericycle.com; if you have distributed the product further, notify your accounts that received the product and ask them to contact Stericycle at 1-866-629-6182 (M-F, 8am-5pm ET) to receive a Response Form. Return affected product to Stericycle using labels provided, if you have not received a return label or reply form or require additional assistance, please contact Stericycle at 1-866-629-6182 (M-F, 8am-5pm) the return labels are for single use only. For additional labels visit http://expertezlabel.com. For further inquiries, please contact ICU Medical Inc. for Global Complaint Management at ProductComplaintsPP@icumed.com or ICU Customer Service at 1-866-829-9025, option 2 (M-F, 8:30am-4pm PT).
Quantity
250 pouches
Distribution
Worldwide Distribution: US (nationwide) to state in AZ, CA, CO, FL, SC, and TX; and countries of: Canada and Japan.
Lot, serial or model codes
Lot number 3483247, UDI(01) 0 0887709 08949 5 (17) 220601 (30) 01 (10) 3483247; lot number 3484722, UDI(01) 0 0887709 08949 5 (17) 220601 (30) 01 (10) 3484722; and lot number 3489328, UDI (01) 0 0887709 08949 5 (17) 220701 (30) 01 (10) 3489328.
510(k) numbers
["K964435"]
PMA numbers
not on file
Product code
FMG
Firm FEI
2025816
Firm city
San Clemente
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register